Open official notice

The Cholestech LDX¿ System combines enzymatic methodology and solidphase technology to measure total cholesterol, HDL cholesterol

Official notice

17 Aug 2022

RecallScope risk estimate:MediumFDA

An official FDA notice concerns The Cholestech LDX¿ System combines enzymatic methodology and solidphase technology to measure total cholesterol, HDL cholesterol. Cholesterol test cassettes package insert is to be updated with new bilirubin interference information because it was found that samples containing bilirubin, at concentrations greater than 2.0 mg/dL (conjugated) or 1.6 mg/dL (unconjugated), can result in low recovery of Triglyceride (TRG) and High-Density Lipoproteins (HDL). Published 2022-08-17.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

436,990

Recall number

Z-1508-2022

Official notice

2022-08-17

The hazard

Cholesterol test cassettes package insert is to be updated with new bilirubin interference information because it was found that samples containing bilirubin, at concentrations greater than 2.0 mg/dL (conjugated) or 1.6 mg/dL (unconjugated), can result in low recovery of Triglyceride (TRG) and High-Density Lipoproteins (HDL).

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

90508

Recall Numbers

Z-1508-2022

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

The Cholestech LDX¿ System combines enzymatic methodology and solidphase technology to measure total cholesterol, HDL cholesterol, triglycerides and glucose. Samples used for testing can be whole blood from a fingerstick (collected in a lithium heparin-coated capillary tube) or venipuncture. The sample is applied to an Cholestech LDX¿ cassette. Description/Part: Cholestech LDX Lipid Profile GLU Cassette/10-991, 99021, 14-531; Cholestech LDX TC HDL GLU Cassette/10-990; Cholestech LDX Lipid Profile Cassette/10-989; Cholestech LDX TC HDL Cassette/10-987

Units affected

436,990

Hazard

Cholesterol test cassettes package insert is to be updated with new bilirubin interference information because it was found that samples containing bilirubin, at concentrations greater than 2.0 mg/dL (conjugated) or 1.6 mg/dL (unconjugated), can result in low recovery of Triglyceride (TRG) and High-Density Lipoproteins (HDL).

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States · AK

Location 2

US · United States · AL

Location 3

US · United States · AR

Location 4

US · United States · AZ

Location 5

US · United States · CA

Location 6

US · United States · CO

Location 7

US · United States · DE

Location 8

US · United States · FL

Location 9

US · United States · GA

Location 10

US · United States · HI

Location 11

US · United States · ID

Location 12

US · United States · IL

Location 13

US · United States · IN

Location 14

US · United States · KS

Location 15

US · United States · KY

Location 16

US · United States · LA

Location 17

US · United States · MA

Location 18

US · United States · MN

Location 19

US · United States · MO

Location 20

US · United States · NC

Location 21

US · United States · NE

Location 22

US · United States · NV

Location 23

US · United States · NY

Location 24

US · United States · OH

Location 25

US · United States · OK

Location 26

US · United States · PA

Location 27

US · United States · RI

Location 28

US · United States · SC

Location 29

US · United States · TN

Location 30

US · United States · TX

Location 31

US · United States · VA

Affected area 1

US · United States

Company & contacts

Manufacturer

Alere San Diego, Inc.

Phone

Not provided

Brand: Alere San Diego. Contact: Alere San Diego, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1508-2022

Event ID

90508

Units affected

436,990

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 17 Aug 2022. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button