TB Series Bipolar Pacing Leads, French Size 4F, 5F, 6F for Curve Type Straight, Atrial J, 60 degree Curve, and Right Heart. TB is packaged

Official notice

14 Nov 2018

RecallScope risk estimate:CriticalFDA

TB Series Bipolar Pacing Leads, French Size 4F, 5F, 6F for Curve Type Straight, Atrial J, 60 degree Curve, and Right Heart. TB is packaged for sterile single use in a double tray. The lead is placed in a PETG inner tray that is sealed with a Tyvek inner lid. The sealed inner tray is placed in an outer PETG tray. The outer tray is sealed with an outer Tyvek lid. An inner label is placed on the outer Tyvek lid. The sealed trays are placed in a labeled white shelf box with on Instruction for use. Packed on lead per outer white shelf box.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

36,688

Recall number

Z-0391-2019

Official notice

2018-11-14

The hazard

During the use of some TB - Temporary Bipolar Pacing Leads, featuring the 2mm unshrouded connectors, the connector cap housing may slide and potentially expose the connection wire. In some instances, this may cause the wire to be more susceptible to loss of connectivity or breakage during movement of the cables causing interruption of the pacing system.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Skus

020004, 020005

Event Ids

80984

Model numbers

020004, 020005

Recall Numbers

Z-0391-2019

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

TB Series Bipolar Pacing Leads, French Size 4F, 5F, 6F for Curve Type Straight, Atrial J, 60 degree Curve, and Right Heart. TB is packaged for sterile single use in a double tray. The lead is placed in a PETG inner tray that is sealed with a Tyvek inner lid. The sealed inner tray is placed in an outer PETG tray. The outer tray is sealed with an outer Tyvek lid. An inner label is placed on the outer Tyvek lid. The sealed trays are placed in a labeled white shelf box with on Instruction for use. Packed on lead per outer white shelf box.

Units affected

36,688

Hazard

During the use of some TB - Temporary Bipolar Pacing Leads, featuring the 2mm unshrouded connectors, the connector cap housing may slide and potentially expose the connection wire. In some instances, this may cause the wire to be more susceptible to loss of connectivity or breakage during movement of the cables causing interruption of the pacing system.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AE · United Arab Emirates

Affected area 3

AT · Austria

Affected area 4

BG · Bulgaria

Affected area 5

BR · Brazil

Affected area 6

CA · Canada

Affected area 7

CO · Colombia

Affected area 8

CR · Costa Rica

Affected area 9

DE · Germany

Affected area 10

DO · Dominican Republic

Affected area 11

EG · Egypt

Affected area 12

GB · United Kingdom

Affected area 13

GR · Greece

Affected area 14

GT · Guatemala

Affected area 15

IL · Israel

Affected area 16

IN · India

Affected area 17

IT · Italy

Affected area 18

JO · Jordan

Affected area 19

KW · Kuwait

Affected area 20

LV · Latvia

Affected area 21

MX · Mexico

Affected area 22

NZ · New Zealand

Affected area 23

PE · Peru

Affected area 24

PK · Pakistan

Affected area 25

PL · Poland

Affected area 26

RU · Russia

Affected area 27

SA · Saudi Arabia

Affected area 28

SV · El Salvador

Affected area 29

TR · Turkey

Affected area 30

TW · Taiwan

Affected area 31

UA · Ukraine

Affected area 32

VE · Venezuela

Affected area 33

VN · Vietnam

Affected area 34

ZA · South Africa

Company & contacts

Brand

Manufacturer

Oscor, Inc.

Phone

Not provided

Brand: Oscor. Contact: Oscor, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class I

Recall details

Recall number

Z-0391-2019

Event ID

80984

Units affected

36,688

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 14 Nov 2018. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button