SynchroMed EL, models 8626-10, 8626L-10, 8626-18, 8626L-18, 8627-10, 8627L-10, 8627-18, 8627L-18. Implantable infusion pump. The pump is
Official notice
19 Dec 2012
An official FDA notice concerns SynchroMed EL, models 8626-10, 8626L-10, 8626-18, 8626L-18, 8627-10, 8627L-10, 8627-18, 8627L-18. Implantable infusion pump. The pump is. Medtronic notified Healthcare Professionals of the impact of unapproved drugs on the performance of the SynchroMed infusion pump system. Use of unapproved drugs with SynchroMed pumps can result in an increased risk of permanent motor stall and cessation of drug infusion. UPDATE 2/1/2016: Medtronic received approval to incorporate enhancements to the SynchroMed II pump which decreases the potential for internal electrical shorting and motor corrosion, and is retrieving any unused pumps. Published 2012-12-19.
What to do now
This recall has been terminated. See the official FDA notice for the latest status.
Units affected
174,700
Recall number
Z-0496-2013
Official notice
2012-12-19
The hazard
The hazard
Medtronic notified Healthcare Professionals of the impact of unapproved drugs on the performance of the SynchroMed infusion pump system. Use of unapproved drugs with SynchroMed pumps can result in an increased risk of permanent motor stall and cessation of drug infusion. UPDATE 2/1/2016: Medtronic received approval to incorporate enhancements to the SynchroMed II pump which decreases the potential for internal electrical shorting and motor corrosion, and is retrieving any unused pumps.
How to check if you’re affected
How to check if you’re affected
Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.
Event Ids
Recall Numbers
Checked manually this time?
Follow brands and keywords in the app to receive alerts for matching official notices.
Turn on recall alertsAffected products
Affected products
Product 1
Units affected
Hazard
Where it was sold
Where it was sold
Retailer information is not currently available in RecallScope. Check the official notice for distribution details.
Location 1
Affected area 1
Affected area 2
Affected area 3
Company & contacts
Company & contacts
Brand: Medtronic Neuromodulation. Contact: Medtronic Neuromodulation.
Source & status
Source & status
Recall details
Recall details
Recall number
Event ID
Units affected
Status
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week’s recalls before you need them.