Open official notice

Stryker, Omega 4.75 mm,. PEEK Knotless Anchor System, Double-Double, REF 3910-500-472, UDI 07613327464832, Sterile R, RX Only CE 0197

Official notice

19 Feb 2020

RecallScope risk estimate:CriticalFDA

Stryker, Omega 4.75 mm,. PEEK Knotless Anchor System, Double-Double, REF 3910-500-472, UDI 07613327464832, Sterile R, Rx Only CE 0197

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

73

Recall number

Z-1147-2020

Official notice

2020-02-19

The hazard

Complaints have been received regarding difficulty in removing the driver from the screw. High driver removal forces can lead to the screw pulling out still attached to the driver. This could result in an additional anchor insertion site in the bone being used.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

84664

Recall Numbers

Z-1147-2020

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Stryker, Omega 4.75 mm,. PEEK Knotless Anchor System, Double-Double, REF 3910-500-472, UDI 07613327464832, Sterile R, RX Only CE 0197

Units affected

73

Hazard

Complaints have been received regarding difficulty in removing the driver from the screw. High driver removal forces can lead to the screw pulling out still attached to the driver. This could result in an additional anchor insertion site in the bone being used.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States · AK

Location 2

US · United States · AR

Location 3

US · United States · AZ

Location 4

US · United States · CA

Location 5

US · United States · CO

Location 6

US · United States · FL

Location 7

US · United States · GA

Location 8

US · United States · ID

Location 9

US · United States · IL

Location 10

US · United States · KY

Location 11

US · United States · MD

Location 12

US · United States · ME

Location 13

US · United States · NJ

Location 14

US · United States · NV

Location 15

US · United States · NY

Location 16

US · United States · OH

Location 17

US · United States · TX

Location 18

US · United States · UT

Location 19

US · United States · WI

Affected area 1

US · United States

Company & contacts

Brand

Manufacturer

Stryker Corporation

Phone

Not provided

Brand: Stryker. Contact: Stryker Corporation.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1147-2020

Event ID

84664

Units affected

73

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 19 Feb 2020. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button