Spectrum IQ Infusion Pump, Product Code 3570009

Official notice

3 Sept 2025

RecallScope risk estimate:CriticalFDA

An official FDA notice concerns Spectrum IQ Infusion Pump, Product Code 3570009. Baxter has identified that certain pumps have potentially been released with the grease applied to the cam and motor gears that will break down quickly. The breakdown of the grease will lead to the device having insufficient or ineffective grease applied to the cam, which could lead to premature wear of the mechanism assembly, resulting in excessive therapy. Published 2025-09-03.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

6

Recall number

Z-2415-2025

Official notice

2025-09-03

The hazard

Baxter has identified that certain pumps have potentially been released with the grease applied to the cam and motor gears that will break down quickly. The breakdown of the grease will lead to the device having insufficient or ineffective grease applied to the cam, which could lead to premature wear of the mechanism assembly, resulting in excessive therapy.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Other

UDI-DI 00085412610900

Event Ids

97335

Recall Numbers

Z-2415-2025

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Spectrum IQ Infusion Pump, Product Code 3570009

Units affected

6

Hazard

Baxter has identified that certain pumps have potentially been released with the grease applied to the cam and motor gears that will break down quickly. The breakdown of the grease will lead to the device having insufficient or ineffective grease applied to the cam, which could lead to premature wear of the mechanism assembly, resulting in excessive therapy.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States · AZ

Location 2

US · United States · FL

Affected area 1

US · United States

Company & contacts

Manufacturer

Baxter Healthcare Corporation

Phone

Not provided

Brand: Baxter Healthcare. Contact: Baxter Healthcare Corporation.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-2415-2025

Event ID

97335

Units affected

6

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 3 Sept 2025. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button