Specialist 2 Intramedullary (SP2 IM) Rod 400mm Instrument (PN 96-6120) The SP2 IM Rod 400mm Instrument (PN 96-6120) is used in both primary

Official notice

14 Oct 2015

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Specialist 2 Intramedullary (SP2 IM) Rod 400mm Instrument (PN 96-6120) The SP2 IM Rod 400mm Instrument (PN 96-6120) is used in both primary. DePuy Orthopaedics, Inc. Is initiating a voluntary recall of select lots of the Specialist 2 Intramedullary Rod 400mm Instrument (PN 96-6120) due to complaints of the rod fracturing and leaving a portion of the rod in patients. Published 2015-10-14.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

7,488

Recall number

Z-0111-2016

Official notice

2015-10-14

The hazard

DePuy Orthopaedics, Inc. Is initiating a voluntary recall of select lots of the Specialist 2 Intramedullary Rod 400mm Instrument (PN 96-6120) due to complaints of the rod fracturing and leaving a portion of the rod in patients.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

s-C98CS4

Event Ids

72113

Recall Numbers

Z-0111-2016

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Specialist 2 Intramedullary (SP2 IM) Rod 400mm Instrument (PN 96-6120) The SP2 IM Rod 400mm Instrument (PN 96-6120) is used in both primary and revision P.F.C.¿ SIGMA¿ knee procedures to align the femoral locating device and distal femoral cutting block. It is also used with the IM tibial resection. This rod is included with the P.F.C.¿ SIGMA¿ Specialist¿ 2 (SP2 / SPII) Primary Knee Instruments, the P.F.C.¿ SIGMA¿ Specialist ¿ 2 (SP2 / SPII) Revision Knee Instruments, and P.F.C. ¿ SIGMA¿ High Performance (HP) Primary Knee Instruments

Units affected

7,488

Hazard

DePuy Orthopaedics, Inc. Is initiating a voluntary recall of select lots of the Specialist 2 Intramedullary Rod 400mm Instrument (PN 96-6120) due to complaints of the rod fracturing and leaving a portion of the rod in patients.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States · AL

Location 2

US · United States · AZ

Location 3

US · United States · CA

Location 4

US · United States · CO

Location 5

US · United States · CT

Location 6

US · United States · DC

Location 7

US · United States · FL

Location 8

US · United States · GA

Location 9

US · United States · HI

Location 10

US · United States · IA

Location 11

US · United States · IL

Location 12

US · United States · IN

Location 13

US · United States · KY

Location 14

US · United States · LA

Location 15

US · United States · MA

Location 16

US · United States · MD

Location 17

US · United States · ME

Location 18

US · United States · MI

Location 19

US · United States · MN

Location 20

US · United States · MO

Location 21

US · United States · MS

Location 22

US · United States · MT

Location 23

US · United States · NC

Location 24

US · United States · NE

Location 25

US · United States · NH

Location 26

US · United States · NJ

Location 27

US · United States · NV

Location 28

US · United States · NY

Location 29

US · United States · OH

Location 30

US · United States · OR

Location 31

US · United States · PA

Location 32

US · United States · RI

Location 33

US · United States · SC

Location 34

US · United States · TN

Location 35

US · United States · TX

Location 36

US · United States · UT

Location 37

US · United States · VA

Location 38

US · United States · VT

Location 39

US · United States · WA

Location 40

US · United States · WI

Location 41

US · United States · WV

Affected area 1

US · United States

Affected area 2

AE · United Arab Emirates

Affected area 3

AR · Argentina

Affected area 4

AT · Austria

Affected area 5

AU · Australia

Affected area 6

BE · Belgium

Affected area 7

BR · Brazil

Affected area 8

CA · Canada

Affected area 9

CH · Switzerland

Affected area 10

CL · Chile

Affected area 11

CN · China

Affected area 12

CO · Colombia

Affected area 13

CZ · Czech Republic

Affected area 14

DE · Germany

Affected area 15

EG · Egypt

Affected area 16

FR · France

Affected area 17

GB · United Kingdom

Affected area 18

GR · Greece

Affected area 19

IL · Israel

Affected area 20

IN · India

Affected area 21

IT · Italy

Affected area 22

KR · South Korea

Affected area 23

MX · Mexico

Affected area 24

NO · Norway

Affected area 25

NZ · New Zealand

Affected area 26

PE · Peru

Affected area 27

PL · Poland

Affected area 28

RU · Russia

Affected area 29

SG · Singapore

Affected area 30

SI · Slovenia

Affected area 31

UY · Uruguay

Affected area 32

VE · Venezuela

Affected area 33

VN · Vietnam

Affected area 34

ZA · South Africa

Company & contacts

Manufacturer

DePuy Orthopaedics, Inc.

Phone

Not provided

Brand: Depuy Orthopaedics. Contact: DePuy Orthopaedics, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0111-2016

Event ID

72113

Units affected

7,488

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 14 Oct 2015. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button