Image not available in RecallScope.
S4C Occiput Torque Wrench F Set Screw; used to tighten the S4 Cervical Spinal System clamping screws with a defined torque 2.8 Nm for
Official notice
12 Aug 2015
An official FDA notice concerns S4C Occiput Torque Wrench F Set Screw; used to tighten the S4 Cervical Spinal System clamping screws with a defined torque 2.8 Nm for. A FW103R torque wrench that was improperly reassembled after reprocessing caused a screw to break during implantation due to too much torque. The firm did not include instructions for reassembly after reprocessing. Published 2015-08-12.
What to do now
This recall has been terminated. See the official FDA notice for the latest status.
Units affected
10
Recall number
Z-2326-2015
Official notice
2015-08-12
The hazard
The hazard
A FW103R torque wrench that was improperly reassembled after reprocessing caused a screw to break during implantation due to too much torque. The firm did not include instructions for reassembly after reprocessing.
How to check if you’re affected
How to check if you’re affected
Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.
Event Ids
Recall Numbers
Checked manually this time?
Follow brands and keywords in the app to receive alerts for matching official notices.
Turn on recall alertsAffected products
Affected products
Product 1
Units affected
Hazard
Where it was sold
Where it was sold
Retailer information is not currently available in RecallScope. Check the official notice for distribution details.
Location 1
Affected area 1
Affected area 2
Company & contacts
Company & contacts
Brand: Aesculap. Contact: Aesculap, Inc..
Source & status
Source & status
Recall details
Recall details
Recall number
Event ID
Units affected
Status
Source tier
Documents, notices & source
See the official notice and source documents for full details.