Open official notice

Rsp standard humeral socket insert, 44mm, hxe-plus, part no. 509-00-044

Official notice

7 Oct 2020

RecallScope risk estimate:MediumFDA

RSP Standard HUMERAL SOCKET INSERT, 44MM, HXe-Plus, Part No. 509-00-044

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

15

Recall number

Z-3021-2020

Official notice

2020-10-07

The hazard

Product may be mislabeled with the size etched onto products differing from the size information shown on patient sticker labels. Risks associated with this issue may include the Surgeon implanting the incorrect size; significant physical impairment, discomfort and/or disruption of treatment of underlying condition; and the patient experiencing dislocation resulting in revision surgery.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

380P1009A

Event Ids

86317

Recall Numbers

Z-3021-2020

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

RSP Standard HUMERAL SOCKET INSERT, 44MM, HXe-Plus, Part No. 509-00-044

Units affected

15

Hazard

Product may be mislabeled with the size etched onto products differing from the size information shown on patient sticker labels. Risks associated with this issue may include the Surgeon implanting the incorrect size; significant physical impairment, discomfort and/or disruption of treatment of underlying condition; and the patient experiencing dislocation resulting in revision surgery.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States · CA

Location 2

US · United States · FL

Location 3

US · United States · GA

Location 4

US · United States · IN

Location 5

US · United States · KS

Location 6

US · United States · MN

Location 7

US · United States · NY

Location 8

US · United States · OK

Location 9

US · United States · PA

Location 10

US · United States · RI

Location 11

US · United States · VA

Location 12

US · United States · WA

Affected area 1

US · United States

Company & contacts

Manufacturer

Encore Medical, LP

Phone

Not provided

Brand: Encore Medical Lp. Contact: Encore Medical, LP.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-3021-2020

Event ID

86317

Units affected

15

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 7 Oct 2020. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.