Open official notice

ResMed Stellar, Non-invasive invasive ventilators, Model: 100, 150

Official notice

12 Feb 2020

RecallScope risk estimate:CriticalFDA

An official FDA notice concerns ResMed Stellar, Non-invasive invasive ventilators, Model: 100, 150. Combination of software and a component failure may cause audible alarms not to operate properly, the alarm buzzer not work, for ventilators that have a failed electronic component and, that are stored without AC power connected for more than 36 hours leading to full depletion of the battery and, that powers on automatically when connected to AC power without pressing the power switch. Published 2020-02-12.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

69

Recall number

Z-0915-2020

Official notice

2020-02-12

The hazard

Combination of software and a component failure may cause audible alarms not to operate properly, the alarm buzzer not work, for ventilators that have a failed electronic component and, that are stored without AC power connected for more than 36 hours leading to full depletion of the battery and, that powers on automatically when connected to AC power without pressing the power switch.

How to check if you're affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

84551

Recall Numbers

Z-0915-2020

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

ResMed Stellar, Non-invasive/invasive ventilators, Model: 100, 150

Units affected

69

Hazard

Combination of software and a component failure may cause audible alarms not to operate properly, the alarm buzzer not work, for ventilators that have a failed electronic component and, that are stored without AC power connected for more than 36 hours leading to full depletion of the battery and, that powers on automatically when connected to AC power without pressing the power switch.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States · CT

Location 2

US · United States · FL

Location 3

US · United States · GA

Location 4

US · United States · IL

Location 5

US · United States · MA

Location 6

US · United States · MT

Location 7

US · United States · PA

Location 8

US · United States · TN

Location 9

US · United States · VA

Affected area 1

US · United States

Company & contacts

Brand

Manufacturer

ResMed Ltd.

Phone

Not provided

Brand: Resmed. Contact: ResMed Ltd..

Source & status

Source

Tier

official

Category

ELECTRONICS

Official classification

Class I

Recall details

Recall number

Z-0915-2020

Event id

84551

Units affected

69

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 12 Feb 2020. The next one won't wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button