Open official notice

Remel thioglycollate medium w o ind. W dextrose w hemin & vit.k (a) 10ml, ref 07182 (b) 10ml, ref 07180 (c) 5ml, ref 064732 (d) 5ml, ref

Official notice

19 Sept 2018

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Remel thioglycollate medium w o ind. W dextrose w hemin & vit.k (a) 10ml, ref 07182 (b) 10ml, ref 07180 (c) 5ml, ref 064732 (d) 5ml, ref. Products have intermittently failed performance testing during stability studies with various anaerobic microorganisms. Published 2018-09-19.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

2,730

Recall number

Z-3017-2018

Official notice

2018-09-19

The hazard

Products have intermittently failed performance testing during stability studies with various anaerobic microorganisms.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

258907, 259526, 271667, 254755, 254749

Event Ids

80813

Recall Numbers

Z-3017-2018

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

remel THIOGLYCOLLATE MEDIUM W/O IND. W/ DEXTROSE W/ HEMIN & VIT.K (a) 10ML, REF 07182 (b) 10ML, REF 07180 (c) 5ML, REF 064732 (d) 5ML, REF 064730 (e) 9ML, REF 064720

Units affected

2,730

Hazard

Products have intermittently failed performance testing during stability studies with various anaerobic microorganisms.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

CA · Canada

Affected area 3

SG · Singapore

Company & contacts

Brand

Manufacturer

Remel Inc

Phone

Not provided

Brand: Remel. Contact: Remel Inc.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-3017-2018

Event ID

80813

Units affected

2,730

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 19 Sept 2018. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button