Reflection(r) dead blow mallet, ref 71362106, qty: (1), non-sterile, smith & nephew, inc,. Memphis, tn 38116

Official notice

18 Jan 2017

RecallScope risk estimate:CriticalFDA

An official FDA notice concerns Reflection(r) dead blow mallet, ref 71362106, qty: (1), non-sterile, smith & nephew, inc,. Memphis, tn 38116. The firm received complaints of cracks in the weld on the head of the mallet. In the reported cases, some of the lead beads and particles escaped from the mallet into the surgical wound. Published 2017-01-18.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

4,444

Recall number

Z-1008-2017

Official notice

2017-01-18

The hazard

The firm received complaints of cracks in the weld on the head of the mallet. In the reported cases, some of the lead beads and particles escaped from the mallet into the surgical wound.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

75751

Recall Numbers

Z-1008-2017

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

REFLECTION(R) DEAD BLOW MALLET, REF 71362106, QTY: (1), NON-STERILE, Smith & Nephew, Inc,. Memphis, TN 38116

Units affected

4,444

Hazard

The firm received complaints of cracks in the weld on the head of the mallet. In the reported cases, some of the lead beads and particles escaped from the mallet into the surgical wound.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AR · Argentina

Affected area 3

AT · Austria

Affected area 4

AU · Australia

Affected area 5

BE · Belgium

Affected area 6

BR · Brazil

Affected area 7

CA · Canada

Affected area 8

CH · Switzerland

Affected area 9

CN · China

Affected area 10

CO · Colombia

Affected area 11

CR · Costa Rica

Affected area 12

DE · Germany

Affected area 13

DK · Denmark

Affected area 14

FI · Finland

Affected area 15

FR · France

Affected area 16

GB · United Kingdom

Affected area 17

GR · Greece

Affected area 18

HK · Hong Kong

Affected area 19

IE · Ireland

Affected area 20

IL · Israel

Affected area 21

IN · India

Affected area 22

IT · Italy

Affected area 23

JP · Japan

Affected area 24

MX · Mexico

Affected area 25

MY · Malaysia

Affected area 26

NL · Netherlands

Affected area 27

NO · Norway

Affected area 28

PL · Poland

Affected area 29

PT · Portugal

Affected area 30

SE · Sweden

Affected area 31

SG · Singapore

Affected area 32

TH · Thailand

Affected area 33

TR · Turkey

Affected area 34

VE · Venezuela

Affected area 35

ZA · South Africa

Company & contacts

Manufacturer

Smith & Nephew, Inc.

Phone

Not provided

Brand: Smith Nephew. Contact: Smith & Nephew, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1008-2017

Event ID

75751

Units affected

4,444

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 18 Jan 2017. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button