Open official notice

R3(tm) 22 mm i.d., 52 mm o.d. Constrained liner, ref 71339152, qty: (1), sterile eo. Product usage: the r3 constrained liner acetabular

Official notice

17 Jun 2015

RecallScope risk estimate:MediumFDA

An official FDA notice concerns R3(tm) 22 mm i.d., 52 mm o.d. Constrained liner, ref 71339152, qty: (1), sterile eo. Product usage: the r3 constrained liner acetabular. The R3 Constrained outer liner locking groove diameter is 0.030 inches oversized and may not allow the locking ring to engage into the R3 shell. Published 2015-06-17.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

17

Recall number

Z-1735-2015

Official notice

2015-06-17

The hazard

The R3 Constrained outer liner locking groove diameter is 0.030 inches oversized and may not allow the locking ring to engage into the R3 shell.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

14HM08235

Event Ids

71330

Recall Numbers

Z-1735-2015

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

R3(TM) 22 MM I.D., 52 MM O.D. CONSTRAINED LINER, REF 71339152, QTY: (1), STERILE EO. Product Usage: The R3 Constrained liner Acetabular System is a cemented or uncemented prosthesis intended to replace a hip joint. The Constrained liner is intended for primary or revision patients at high risk for hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability and for whom all other options to constrained acetabular components have been considered. The R3 Constrained liner is intended for single use only.

Units affected

17

Hazard

The R3 Constrained outer liner locking groove diameter is 0.030 inches oversized and may not allow the locking ring to engage into the R3 shell.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AU · Australia

Affected area 3

CA · Canada

Affected area 4

CH · Switzerland

Affected area 5

CL · Chile

Affected area 6

ES · Spain

Company & contacts

Manufacturer

Smith & Nephew, Inc.

Phone

Not provided

Brand: Smith Nephew. Contact: Smith & Nephew, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1735-2015

Event ID

71330

Units affected

17

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 17 Jun 2015. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.