Open official notice

Quickslide GramPro 1 Automated Gram Stanier is a microscope slide stainer

Official notice

16 Jun 2021

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Quickslide GramPro 1 Automated Gram Stanier is a microscope slide stainer. Instruments were assembled and serviced using a defective lot of tubing, which is part of the pump mechanism within the unit. Published 2021-06-16.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

69

Recall number

Z-1835-2021

Official notice

2021-06-16

The hazard

Instruments were assembled and serviced using a defective lot of tubing, which is part of the pump mechanism within the unit.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

1654

Event Ids

87850

Recall Numbers

Z-1835-2021

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Quickslide GramPro 1 Automated Gram Stanier is a microscope slide stainer.

Units affected

69

Hazard

Instruments were assembled and serviced using a defective lot of tubing, which is part of the pump mechanism within the unit.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Company & contacts

Manufacturer

Hardy Diagnostics

Phone

Not provided

Brand: Hardy Diagnostics. Contact: Hardy Diagnostics.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1835-2021

Event ID

87850

Units affected

69

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 16 Jun 2021. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button