Open official notice

Pt113950 pt hybrid glenoid post regenerex porous titanium construct sterile r prosthesis, shoulder, semi-constrained, metal polymer cemented

Official notice

16 Jan 2013

RecallScope risk estimate:CriticalFDA

An official FDA notice concerns Pt113950 pt hybrid glenoid post regenerex porous titanium construct sterile r prosthesis, shoulder, semi-constrained, metal polymer cemented. Biomet is recalling Part Number PT-113950 PT Hybrid Glenoid Post, following an investigation which identified that the male thread of the post may be oversized. This oversized condition can vary in degree and may lead to the following three events: 1) If the PT Hybrid Glenoid Post is not fully seated into the Hybrid Base and the implant construct is implanted, then a gap of 1-3 mm will be prese Published 2013-01-16.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

1,050

Recall number

Z-0660-2013

Official notice

2013-01-16

The hazard

Biomet is recalling Part Number PT-113950 PT Hybrid Glenoid Post, following an investigation which identified that the male thread of the post may be oversized. This oversized condition can vary in degree and may lead to the following three events: 1) If the PT Hybrid Glenoid Post is not fully seated into the Hybrid Base and the implant construct is implanted, then a gap of 1-3 mm will be prese

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

093540

Event Ids

63850

Recall Numbers

Z-0660-2013

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

PT113950 PT HYBRID Glenoid Post Regenerex porous titanium construct Sterile R Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented

Units affected

1,050

Hazard

Biomet is recalling Part Number PT-113950 PT Hybrid Glenoid Post, following an investigation which identified that the male thread of the post may be oversized. This oversized condition can vary in degree and may lead to the following three events: 1) If the PT Hybrid Glenoid Post is not fully seated into the Hybrid Base and the implant construct is implanted, then a gap of 1-3 mm will be prese

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AR · Argentina

Affected area 3

AU · Australia

Affected area 4

CA · Canada

Affected area 5

CL · Chile

Affected area 6

JP · Japan

Affected area 7

NL · Netherlands

Company & contacts

Brand

Manufacturer

Biomet, Inc.

Phone

Not provided

Brand: Biomet. Contact: Biomet, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0660-2013

Event ID

63850

Units affected

1,050

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 16 Jan 2013. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button