An official FDA notice concerns Prosthesis, knee, patello femorotibial, semi-constrained, uncemented, porous, coated, polymer metal polymer. Stryker has discovered that the internal profile of the Triathlon Tritanium Tibial Component (Baseplate), size 4, is potentially oversized. Due to a manufacturing issue in Lot Number CTD44301, there may be a witness mark (taper) at the bottom of the internal profile of the baseplate. The taper prevents the insert from locking on the baseplate. Published 2021-03-17.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

11

Recall number

Z-1186-2021

Official notice

2021-03-17

The hazard

Stryker has discovered that the internal profile of the Triathlon Tritanium Tibial Component (Baseplate), size 4, is potentially oversized. Due to a manufacturing issue in Lot Number CTD44301, there may be a witness mark (taper) at the bottom of the internal profile of the baseplate. The taper prevents the insert from locking on the baseplate.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

CTD44301

Event Ids

87350

Recall Numbers

Z-1186-2021

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Prosthesis, knee, patello/femorotibial, semi-constrained, uncemented, porous, coated, polymer/metal/polymer.

Units affected

11

Hazard

Stryker has discovered that the internal profile of the Triathlon Tritanium Tibial Component (Baseplate), size 4, is potentially oversized. Due to a manufacturing issue in Lot Number CTD44301, there may be a witness mark (taper) at the bottom of the internal profile of the baseplate. The taper prevents the insert from locking on the baseplate.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States · AZ

Location 2

US · United States · GA

Location 3

US · United States · MA

Location 4

US · United States · MI

Location 5

US · United States · MN

Location 6

US · United States · NH

Location 7

US · United States · NJ

Location 8

US · United States · OR

Location 9

US · United States · PA

Location 10

US · United States · SC

Location 11

US · United States · UT

Location 12

US · United States · WA

Location 13

US · United States · WV

Affected area 1

US · United States

Company & contacts

Manufacturer

Howmedica Osteonics Corp.

Phone

Not provided

Brand: Howmedica Osteonics. Contact: Howmedica Osteonics Corp..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1186-2021

Event ID

87350

Units affected

11

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 17 Mar 2021. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button