Open official notice

Pre-filled Water Syringe, Model Numbers SW-010-100 and SW-010

Official notice

11 Jun 2025

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Pre-filled Water Syringe, Model Numbers SW-010-100 and SW-010. During endoscope surveillance monitoring, a user facility identified contamination from duodenoscope samples that were linked to the Pre-filled Water Syringes as part of their investigation. Healthmark Industries conducted its own investigation and concluded that the Pre-filled Water Syringe had been improperly used outside of the intended purpose. Published 2025-06-11.

What to do now

Do not consume or use the affected product. Return it to the place of purchase or follow instructions in the official FDA notice.

Units affected

250,000

Recall number

Z-1909-2025

Official notice

2025-06-11

The hazard

During endoscope surveillance monitoring, a user facility identified contamination from duodenoscope samples that were linked to the Pre-filled Water Syringes as part of their investigation. Healthmark Industries conducted its own investigation and concluded that the Pre-filled Water Syringe had been improperly used outside of the intended purpose.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

200001

Event Ids

96823

Recall Numbers

Z-1909-2025

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Pre-filled Water Syringe, Model Numbers SW-010-100 and SW-010

Units affected

250,000

Hazard

During endoscope surveillance monitoring, a user facility identified contamination from duodenoscope samples that were linked to the Pre-filled Water Syringes as part of their investigation. Healthmark Industries conducted its own investigation and concluded that the Pre-filled Water Syringe had been improperly used outside of the intended purpose.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

CA · Canada

Affected area 3

MY · Malaysia

Company & contacts

Manufacturer

Healthmark Industries Co., Inc.

Phone

Not provided

Brand: Healthmark. Contact: Healthmark Industries Co., Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1909-2025

Event ID

96823

Units affected

250,000

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 11 Jun 2025. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.