Open official notice

Philips CT systems labeled as the following: 1. Brilliance CT 40 Channel, Model Number: 728235. 2. Brilliance CT 64 Channel, Model Number

Official notice

1 Oct 2025

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Philips CT systems labeled as the following: 1. Brilliance CT 40 Channel, Model Number: 728235. 2. Brilliance CT 64 Channel, Model Number. The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement. Published 2025-10-01.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

181

Recall number

Z-2595-2025

Official notice

2025-10-01

The hazard

The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Skus

728235, 728231, 728317

Event Ids

97468

Model numbers

728235, 728231, 728317

Recall Numbers

Z-2595-2025

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Philips CT systems labeled as the following: 1. Brilliance CT 40 Channel, Model Number: 728235. 2. Brilliance CT 64 Channel, Model Number: 728231. 3. Ingenuity Flex, Model Number: 728317

Units affected

181

Hazard

The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AR · Argentina

Affected area 3

AT · Austria

Affected area 4

AU · Australia

Affected area 5

BE · Belgium

Affected area 6

BR · Brazil

Affected area 7

CA · Canada

Affected area 8

CH · Switzerland

Affected area 9

CL · Chile

Affected area 10

CN · China

Affected area 11

CZ · Czech Republic

Affected area 12

DE · Germany

Affected area 13

DK · Denmark

Affected area 14

EG · Egypt

Affected area 15

ES · Spain

Affected area 16

FR · France

Affected area 17

GB · United Kingdom

Affected area 18

GR · Greece

Affected area 19

HK · Hong Kong

Affected area 20

HU · Hungary

Affected area 21

IE · Ireland

Affected area 22

IL · Israel

Affected area 23

IN · India

Affected area 24

IT · Italy

Affected area 25

JO · Jordan

Affected area 26

JP · Japan

Affected area 27

KR · South Korea

Affected area 28

LV · Latvia

Affected area 29

MX · Mexico

Affected area 30

NL · Netherlands

Affected area 31

NO · Norway

Affected area 32

PE · Peru

Affected area 33

PL · Poland

Affected area 34

PT · Portugal

Affected area 35

RO · Romania

Affected area 36

SE · Sweden

Affected area 37

TH · Thailand

Affected area 38

TW · Taiwan

Affected area 39

VN · Vietnam

Company & contacts

Brand

Manufacturer

Philips North America Llc

Phone

Not provided

Brand: Philips. Contact: Philips North America Llc.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-2595-2025

Event ID

97468

Units affected

181

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 1 Oct 2025. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button