Open official notice

Persona Partial Knee spacer block instruments; (1) 8mm block, Model Number: 42-5399-035-08; (2) 9mm block, Model Number:42-5399-035-09; (3)

Official notice

4 Apr 2018

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Persona Partial Knee spacer block instruments; (1) 8mm block, Model Number: 42-5399-035-08; (2) 9mm block, Model Number:42-5399-035-09; (3). Certain partial knee implant spacer blocks do not fit in the alignment tower instrument or handle. Published 2018-04-04.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

1,023

Recall number

Z-1238-2018

Official notice

2018-04-04

The hazard

Certain partial knee implant spacer blocks do not fit in the alignment tower instrument or handle.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Skus

42-5399-035-08, 42-5399-035-09, 42-5399-035-10, 42-5399-035-12, 42-5399-035-14

Event Ids

79383

Model numbers

42-5399-035-08, 42-5399-035-09, 42-5399-035-10, 42-5399-035-12, 42-5399-035-14

Recall Numbers

Z-1238-2018

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Persona Partial Knee spacer block instruments; (1) 8mm block, Model Number: 42-5399-035-08; (2) 9mm block, Model Number:42-5399-035-09; (3) 10mm block, Model Number: 42-5399-035-10; (4) 12mm block, Model Number: 42-5399-035-12; (5) 14mm block, Model Number: 42-5399-035-14 A spacer block is used to ensure the tibial resection Is sufficient and to verify limb alignment. After verification the distal femoral resecUon Is complete using the distal femoral cut slot in the spacer block. The spacer block determines how much distal femoral bone is resected.

Units affected

1,023

Hazard

Certain partial knee implant spacer blocks do not fit in the alignment tower instrument or handle.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

JP · Japan

Affected area 3

NL · Netherlands

Company & contacts

Manufacturer

Zimmer Biomet, Inc.

Phone

Not provided

Brand: Zimmer Biomet. Contact: Zimmer Biomet, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1238-2018

Event ID

79383

Units affected

1,023

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 4 Apr 2018. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button