Open official notice

Persona Cemented Tibial Broach Inserter Extractor. The Persona Cemented Tibial Broach Inserter Extractor Handle is a Use sterilizable

Official notice

31 Dec 2014

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Persona Cemented Tibial Broach Inserter Extractor. The Persona Cemented Tibial Broach Inserter Extractor Handle is a Use sterilizable. The affected lot was supplied by contract manufacturer and are potentially subject to disassociation of the magnet cover. Zimmer received one complaint indicating the magnet cover of a Persona Cemented Tibial Broach Inserter/Extractor Handle lot 62085420 was missing after broaching the tibia. Postoperative radiographs were inspected and the magnet cover was not identified. The investigation indi Published 2014-12-31.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

83

Recall number

Z-0853-2015

Official notice

2014-12-31

The hazard

The affected lot was supplied by contract manufacturer and are potentially subject to disassociation of the magnet cover. Zimmer received one complaint indicating the magnet cover of a Persona Cemented Tibial Broach Inserter/Extractor Handle lot 62085420 was missing after broaching the tibia. Postoperative radiographs were inspected and the magnet cover was not identified. The investigation indi

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

62085420

Event Ids

69906

Recall Numbers

Z-0853-2015

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Persona Cemented Tibial Broach Inserter/Extractor. The Persona Cemented Tibial Broach Inserter/Extractor Handle is a Use sterilizable instrument intended for use In multiple TKA procedures. The Persona Surgical Technique indicates that the Persona Cemented Tibial Broach Inserter/Extractor Handle is intended to broach the proximal tibia in preparation for the full seating of the tibial implant.

Units affected

83

Hazard

The affected lot was supplied by contract manufacturer and are potentially subject to disassociation of the magnet cover. Zimmer received one complaint indicating the magnet cover of a Persona Cemented Tibial Broach Inserter/Extractor Handle lot 62085420 was missing after broaching the tibia. Postoperative radiographs were inspected and the magnet cover was not identified. The investigation indi

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States · AL

Location 2

US · United States · CA

Location 3

US · United States · CO

Location 4

US · United States · FL

Location 5

US · United States · IL

Location 6

US · United States · MI

Location 7

US · United States · MN

Location 8

US · United States · NC

Location 9

US · United States · NJ

Location 10

US · United States · OH

Location 11

US · United States · OR

Location 12

US · United States · PA

Affected area 1

US · United States

Affected area 2

AE · United Arab Emirates

Affected area 3

AU · Australia

Affected area 4

DE · Germany

Affected area 5

IN · India

Affected area 6

IT · Italy

Company & contacts

Brand

Manufacturer

Zimmer, Inc.

Phone

Not provided

Brand: Zimmer. Contact: Zimmer, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0853-2015

Event ID

69906

Units affected

83

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 31 Dec 2014. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button