Open official notice

PENTAX Medical Video Colonoscope- Intended to provide optical visualization of (via a video monitor), and therapeutic access to the lower

Official notice

5 Mar 2025

RecallScope risk estimate:MediumFDA

PENTAX Medical Video Colonoscope- Intended to provide optical visualization of (via a video monitor), and therapeutic access to the lower gastrointestinal tract Model: EC38-i20cL

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

152

Recall number

Z-1236-2025

Official notice

2025-03-05

The hazard

During endoscopic procedures using a combination of the video processor EPK-i8020c and the i20c series video endoscope, the observed image can become reddish or dark (1.8%). Users have observed smoke-like steam and noted that the light guide at the tip is hot during use / after the endoscope is removed from the patient, and may cause thermal injury to the patient's mucous membranes.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

96164

Recall Numbers

Z-1236-2025

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

PENTAX Medical Video Colonoscope- Intended to provide optical visualization of (via a video monitor), and therapeutic access to the lower gastrointestinal tract Model: EC38-i20cL

Units affected

152

Hazard

During endoscopic procedures using a combination of the video processor EPK-i8020c and the i20c series video endoscope, the observed image can become reddish or dark (1.8%). Users have observed smoke-like steam and noted that the light guide at the tip is hot during use / after the endoscope is removed from the patient, and may cause thermal injury to the patient's mucous membranes.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Company & contacts

Manufacturer

Pentax of America Inc

Phone

Not provided

Brand: Pentax Of. Contact: Pentax of America Inc.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1236-2025

Event ID

96164

Units affected

152

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 5 Mar 2025. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button