Open official notice

Pds ii (polydioxanone) suture- violet monofilament 27in 0 s a ct-1 product code: z340h for use in soft tissue approximation, including use

Official notice

21 Dec 2022

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Pds ii (polydioxanone) suture- violet monofilament 27in 0 s a ct-1 product code: z340h for use in soft tissue approximation, including use. Knot tensile strengths were observed to be significantly below the minimum knot tensile, will result in breakage prior to use or intra-operatively rather than premature post-operative breakage Published 2022-12-21.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

7,560

Recall number

Z-0480-2023

Official notice

2022-12-21

The hazard

Knot tensile strengths were observed to be significantly below the minimum knot tensile, will result in breakage prior to use or intra-operatively rather than premature post-operative breakage

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Other

UDI-DI 10705031060623

Event Ids

91192

Recall Numbers

Z-0480-2023

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

PDS II (polydioxanone) Suture- VIOLET MONOFILAMENT 27IN 0 S/A CT-1 Product Code: Z340H For use in soft tissue approximation, including use in pediatric cardiovascular tissue where growth is expected to occur and ophthalmic surgery

Units affected

7,560

Hazard

Knot tensile strengths were observed to be significantly below the minimum knot tensile, will result in breakage prior to use or intra-operatively rather than premature post-operative breakage

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AU · Australia

Affected area 3

CA · Canada

Affected area 4

JP · Japan

Affected area 5

NZ · New Zealand

Company & contacts

Brand

Manufacturer

Ethicon, Inc.

Phone

Not provided

Brand: Ethicon. Contact: Ethicon, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0480-2023

Event ID

91192

Units affected

7,560

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 21 Dec 2022. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button