Open official notice

Oxoid Nitrocefin (+ Reconstitution Fluid Batch) REF: SR0112C, Oxoid Ltd. For the rapid chromogenic detection of beta-lactamase activity in

Official notice

15 Jun 2016

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Oxoid Nitrocefin (+ Reconstitution Fluid Batch) REF: SR0112C, Oxoid Ltd. For the rapid chromogenic detection of beta-lactamase activity in. Individual bottles of product may produce weak beta-lactamase reactions. Use may result in delayed or false negative reporting. Published 2016-06-15.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

2

Recall number

Z-1921-2016

Official notice

2016-06-15

The hazard

Individual bottles of product may produce weak beta-lactamase reactions. Use may result in delayed or false negative reporting.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

1717357

Event Ids

73658

Recall Numbers

Z-1921-2016

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Oxoid Nitrocefin (+ Reconstitution Fluid Batch) REF: SR0112C, Oxoid Ltd. For the rapid chromogenic detection of beta-lactamase activity in bacterial cells. in vitro diagnostic

Units affected

2

Hazard

Individual bottles of product may produce weak beta-lactamase reactions. Use may result in delayed or false negative reporting.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States · MO

Affected area 1

US · United States

Company & contacts

Brand

Manufacturer

Remel Inc

Phone

Not provided

Brand: Remel. Contact: Remel Inc.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1921-2016

Event ID

73658

Units affected

2

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 15 Jun 2016. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button