Open official notice

Oncomine Dx Target Test, (Model No. A32451), Oncomine Dx Target RNA DNA Panel (Model No. A32441). A qualitative In Vitro Diagnostic test

Official notice

2 May 2018

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Oncomine Dx Target Test, (Model No. A32451), Oncomine Dx Target RNA DNA Panel (Model No. A32441). A qualitative In Vitro Diagnostic test. The lower concentration of the RNA panel could cause a NO CALL or FALSE NEGATIVE for the reporting of ROS1 fusion. If a physician relies on a false negative result, crizotinib could be improperly excluded from the patient's treatment regimen. Published 2018-05-02.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

30

Recall number

Z-1480-2018

Official notice

2018-05-02

The hazard

The lower concentration of the RNA panel could cause a NO CALL or FALSE NEGATIVE for the reporting of ROS1 fusion. If a physician relies on a false negative result, crizotinib could be improperly excluded from the patient's treatment regimen.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

1705001

Event Ids

79649

Recall Numbers

Z-1480-2018

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Oncomine Dx Target Test, (Model No. A32451), Oncomine Dx Target RNA/DNA Panel (Model No. A32441). A qualitative In Vitro Diagnostic test.

Units affected

30

Hazard

The lower concentration of the RNA panel could cause a NO CALL or FALSE NEGATIVE for the reporting of ROS1 fusion. If a physician relies on a false negative result, crizotinib could be improperly excluded from the patient's treatment regimen.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States · AZ

Location 2

US · United States · CA

Location 3

US · United States · NC

Location 4

US · United States · TX

Location 5

US · United States · VA

Affected area 1

US · United States

Company & contacts

Brand

Manufacturer

Life Technologies Corporation

Phone

Not provided

Brand: Life. Contact: Life Technologies Corporation.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1480-2018

Event ID

79649

Units affected

30

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 2 May 2018. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button