Open official notice

Olympic Cool-Cap System is intended to provide treatment for neonatal hypoxic-ischemic encephalopathy (HIE). The Cool-Cap is indicated for

Official notice

27 Jun 2012

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Olympic Cool-Cap System is intended to provide treatment for neonatal hypoxic-ischemic encephalopathy (HIE). The Cool-Cap is indicated for. Olympic Cool-Cap Cooling Module's power supply failed during treatment. The power supply failure can happen without prior indication and results in the unavailability of the system to begin or continue cooling treatment. Published 2012-06-27.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

181

Recall number

Z-1843-2012

Official notice

2012-06-27

The hazard

Olympic Cool-Cap Cooling Module's power supply failed during treatment. The power supply failure can happen without prior indication and results in the unavailability of the system to begin or continue cooling treatment.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

61946

Recall Numbers

Z-1843-2012

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Olympic Cool-Cap System is intended to provide treatment for neonatal hypoxic-ischemic encephalopathy (HIE). The Cool-Cap is indicated for use in full term infants (more than 36 week gestational age) with clinical evidence of moderate to severe HIE. The device provides selective head cooling with mild systemic hypothermia to prevent or reduce the severity of neurologic injury associated with HIE. It is a selective head cooling system for treatment of neonatal hypoxic-ischemic encephalopathy (HIE) in infants. A cooling water cap is placed on the infant's head and the body is warmed using radiant warmers. The Cool-Cap provides 72 hours of monitored cooling. The device is labeled in part: "***OLYMPIC COOL-CAP Cooling Unit***Model No.***Made in USA***Software in this product is Natus Medical Incorporated***"

Units affected

181

Hazard

Olympic Cool-Cap Cooling Module's power supply failed during treatment. The power supply failure can happen without prior indication and results in the unavailability of the system to begin or continue cooling treatment.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AE · United Arab Emirates

Affected area 3

AR · Argentina

Affected area 4

AT · Austria

Affected area 5

AU · Australia

Affected area 6

CA · Canada

Affected area 7

DE · Germany

Affected area 8

EG · Egypt

Affected area 9

ES · Spain

Affected area 10

GB · United Kingdom

Affected area 11

HK · Hong Kong

Affected area 12

IT · Italy

Affected area 13

KR · South Korea

Affected area 14

KW · Kuwait

Affected area 15

PL · Poland

Affected area 16

RU · Russia

Affected area 17

SA · Saudi Arabia

Affected area 18

SG · Singapore

Affected area 19

SI · Slovenia

Affected area 20

TH · Thailand

Affected area 21

TR · Turkey

Affected area 22

TW · Taiwan

Affected area 23

ZA · South Africa

Company & contacts

Manufacturer

Natus Medical Incorporated

Phone

Not provided

Brand: Natus Medical. Contact: Natus Medical Incorporated.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1843-2012

Event ID

61946

Units affected

181

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 27 Jun 2012. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button