Open official notice

NanoTips 30g 20pk, 100pk PN 1240, 1426 Description: NanoTips are injection molded royal blue tips that contain stainless steel 30-gauge

Official notice

25 Jan 2017

RecallScope risk estimate:MediumFDA

NanoTips 30g 20pk, 100pk PN 1240, 1426 Description: NanoTips are injection molded royal blue tips that contain stainless steel 30-gauge cannula to aid in the delivery of chemicals. The NanoTip base is made from a polypropylene copolymer. The stainless steel cannula is made from 304 SS, and is 30-gauge in diameter and 0.488 in length. The cannula is secured in the base with an epoxy adhesive. Indications: Tips, when used in conjunction with the range of available syringes and products, provide a technique that is designed to provide a faster, easier, and more precise delivery. They are designed and created in such a way to generate less product waste and provide optimum control/leverage. Tips are also used for moisture removal and for directing laser light beam. Each of the tips is part of delivery systems designed for dentist-supervised, intra-oral procedures. The NanoTIp is designed for use with dental products manufactured by Ultradent Products, Inc. Such as: UltraSeal XT¿ plus, UltraSeal XT¿ hydro, Ultra-Etch¿, PermaFloIdeal for microscopic restorative dentistry

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

143

Recall number

Z-1034-2017

Official notice

2017-01-25

The hazard

Ultradent Products, Inc. Announces a voluntary field action for the NanoTips 30g 20pk, 100pk because the cannula attached to the tip hub may become loose or detach from the blue base that connects to the syringe.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

BD6KM

Event Ids

76039

Recall Numbers

Z-1034-2017

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

NanoTips 30g 20pk, 100pk PN 1240, 1426 Description: NanoTips are injection molded royal blue tips that contain stainless steel 30-gauge cannula to aid in the delivery of chemicals. The NanoTip base is made from a polypropylene copolymer. The stainless steel cannula is made from 304 SS, and is 30-gauge in diameter and 0.488 in length. The cannula is secured in the base with an epoxy adhesive. Indications: Tips, when used in conjunction with the range of available syringes and products, provide a technique that is designed to provide a faster, easier, and more precise delivery. They are designed and created in such a way to generate less product waste and provide optimum control/leverage. Tips are also used for moisture removal and for directing laser light beam. Each of the tips is part of delivery systems designed for dentist-supervised, intra-oral procedures. The NanoTIp is designed for use with dental products manufactured by Ultradent Products, Inc. such as: UltraSeal XT¿ plus, UltraSeal XT¿ hydro, Ultra-Etch¿, PermaFloIdeal for microscopic restorative dentistry

Units affected

143

Hazard

Ultradent Products, Inc. Announces a voluntary field action for the NanoTips 30g 20pk, 100pk because the cannula attached to the tip hub may become loose or detach from the blue base that connects to the syringe.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

BE · Belgium

Affected area 3

CH · Switzerland

Affected area 4

DE · Germany

Affected area 5

IT · Italy

Affected area 6

JP · Japan

Affected area 7

KR · South Korea

Affected area 8

NL · Netherlands

Company & contacts

Manufacturer

Ultradent Products, Inc.

Phone

Not provided

Brand: Ultradent. Contact: Ultradent Products, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1034-2017

Event ID

76039

Units affected

143

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 25 Jan 2017. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button