Merge Hemo with software versions 10.0 LA. Merge Hemo (formerly named HeartSuite Hemodynamics) monitors, measures, and records physiologic data from a human patient undergoing a cardiac catheterization procedure.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

2

Recall number

Z-0554-2016

Official notice

2016-01-06

The hazard

The firm, Merge Healthcare, sent out Merge HEMO V10.0 & Merge CARDIO V10.1 software to 3 customers before the product was completely validated in house.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

72661

Recall Numbers

Z-0554-2016

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Merge Hemo with software versions 10.0 LA. Merge Hemo (formerly named HeartSuite Hemodynamics) monitors, measures, and records physiologic data from a human patient undergoing a cardiac catheterization procedure.

Units affected

2

Hazard

The firm, Merge Healthcare, sent out Merge HEMO V10.0 & Merge CARDIO V10.1 software to 3 customers before the product was completely validated in house.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States · IL

Location 2

US · United States · NC

Location 3

US · United States · VT

Affected area 1

US · United States

Company & contacts

Manufacturer

Merge Healthcare, Inc.

Phone

Not provided

Brand: Merge Healthcare. Contact: Merge Healthcare, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0554-2016

Event ID

72661

Units affected

2

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 6 Jan 2016. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button