Open official notice

Merge, eFilm Workstation and eFilm Lite eFilm Workstation with Modules is a software application that is used for viewing medical images

Official notice

14 Dec 2016

RecallScope risk estimate:MediumFDA

Merge, eFilm Workstation and eFilm Lite eFilm Workstation with Modules is a software application that is used for viewing medical images. EFilm Workstation with Modules receives digital images and data from various sources (including but not limited to CT, MR, US, RF units, computed and direct radiographic devices, secondary capture devices, scanners, imaging gateways or imaging sources). EFilm Workstation with Modules can be used to communicate, process and display medical images. Users have access to various image processing and measurement tools to assist them in viewing images. In addition, users can overlay templates on medical images to aid in preoperative planning. EFilm Workstation with Modules can be integrated with an institution s existing HIS or RIS for a fully integrated electronic patient record. Typical users of eFilm Workstation with Modules are trained medical professionals, including but not limited to radiologists, technologists and clinicians

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

2,163

Recall number

Z-0707-2017

Official notice

2016-12-14

The hazard

A product issues happens for RF projection images. If user measures on RF image, .cal (Calibration) is not displayed and measurement is not correct. The .cal label is not displayed on the projection images after calibration is performed which should be present if the measurement is presented based on Pixel Spacing.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

75270

Recall Numbers

Z-0707-2017

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Merge, eFilm Workstation and eFilm Lite eFilm Workstation with Modules is a software application that is used for viewing medical images. eFilm Workstation with Modules receives digital images and data from various sources (including but not limited to CT, MR, US, RF units, computed and direct radiographic devices, secondary capture devices, scanners, imaging gateways or imaging sources). eFilm Workstation with Modules can be used to communicate, process and display medical images. Users have access to various image processing and measurement tools to assist them in viewing images. In addition, users can overlay templates on medical images to aid in preoperative planning. eFilm Workstation with Modules can be integrated with an institution s existing HIS or RIS for a fully integrated electronic patient record. Typical users of eFilm Workstation with Modules are trained medical professionals, including but not limited to radiologists, technologists and clinicians

Units affected

2,163

Hazard

A product issues happens for RF projection images. If user measures on RF image, .cal (Calibration) is not displayed and measurement is not correct. The .cal label is not displayed on the projection images after calibration is performed which should be present if the measurement is presented based on Pixel Spacing.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AE · United Arab Emirates

Affected area 3

AR · Argentina

Affected area 4

AT · Austria

Affected area 5

AU · Australia

Affected area 6

BE · Belgium

Affected area 7

BR · Brazil

Affected area 8

CA · Canada

Affected area 9

CH · Switzerland

Affected area 10

CL · Chile

Affected area 11

CN · China

Affected area 12

CO · Colombia

Affected area 13

CR · Costa Rica

Affected area 14

CY · Cyprus

Affected area 15

CZ · Czech Republic

Affected area 16

DE · Germany

Affected area 17

DK · Denmark

Affected area 18

EE · Estonia

Affected area 19

ES · Spain

Affected area 20

FI · Finland

Affected area 21

FR · France

Affected area 22

GB · United Kingdom

Affected area 23

GR · Greece

Affected area 24

GT · Guatemala

Affected area 25

HK · Hong Kong

Affected area 26

HR · Croatia

Affected area 27

HU · Hungary

Affected area 28

IE · Ireland

Affected area 29

IL · Israel

Affected area 30

IN · India

Affected area 31

IS · Iceland

Affected area 32

IT · Italy

Affected area 33

JP · Japan

Affected area 34

LT · Lithuania

Affected area 35

LU · Luxembourg

Affected area 36

LV · Latvia

Affected area 37

MX · Mexico

Affected area 38

MY · Malaysia

Affected area 39

NL · Netherlands

Affected area 40

NO · Norway

Affected area 41

NZ · New Zealand

Affected area 42

PE · Peru

Affected area 43

PK · Pakistan

Affected area 44

PL · Poland

Affected area 45

PT · Portugal

Affected area 46

RO · Romania

Affected area 47

RS · Serbia

Affected area 48

RU · Russia

Affected area 49

SA · Saudi Arabia

Affected area 50

SE · Sweden

Affected area 51

SG · Singapore

Affected area 52

SI · Slovenia

Affected area 53

SK · Slovakia

Affected area 54

TR · Turkey

Affected area 55

UA · Ukraine

Affected area 56

UY · Uruguay

Affected area 57

VN · Vietnam

Affected area 58

ZA · South Africa

Company & contacts

Manufacturer

Merge Healthcare, Inc.

Phone

Not provided

Brand: Merge Healthcare. Contact: Merge Healthcare, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0707-2017

Event ID

75270

Units affected

2,163

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 14 Dec 2016. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button