Open official notice

Medtronic Nexdrive Micropositioning Drive. Models Ml-1000 and Ml-2000. For use in conjunction with the Medtronic Nexframe Stereotactic

Official notice

6 Aug 2014

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Medtronic Nexdrive Micropositioning Drive. Models Ml-1000 and Ml-2000. For use in conjunction with the Medtronic Nexframe Stereotactic. Potential for misalignment of the Z-stage scale. Using one of these devices for a procedure could result in the microelectrode being inserted to an incorrect target depth. Published 2014-08-06.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

300

Recall number

Z-2118-2014

Official notice

2014-08-06

The hazard

Potential for misalignment of the Z-stage scale. Using one of these devices for a procedure could result in the microelectrode being inserted to an incorrect target depth.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

082605614, 082616214

Event Ids

68615

Recall Numbers

Z-2118-2014

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Medtronic Nexdrive Micropositioning Drive. Models Ml-1000 and Ml-2000. For use in conjunction with the Medtronic Nexframe Stereotactic System for the precise positioning of microelectrodes and implantable leads. A stereotactic guidance system used in conjunction with Medtronic StealthStation Navigation Systems-image-guided surgery (IGS) systems-for Deep Brain Stimulation (DBS) procedures.

Units affected

300

Hazard

Potential for misalignment of the Z-stage scale. Using one of these devices for a procedure could result in the microelectrode being inserted to an incorrect target depth.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States · CA

Location 2

US · United States · CO

Location 3

US · United States · FL

Location 4

US · United States · IL

Location 5

US · United States · KS

Location 6

US · United States · KY

Location 7

US · United States · LA

Location 8

US · United States · MN

Location 9

US · United States · MT

Location 10

US · United States · NC

Location 11

US · United States · NE

Location 12

US · United States · OR

Location 13

US · United States · TX

Location 14

US · United States · UT

Location 15

US · United States · VA

Affected area 1

US · United States

Affected area 2

CA · Canada

Affected area 3

ES · Spain

Affected area 4

GB · United Kingdom

Company & contacts

Manufacturer

Medtronic Neuromodulation

Phone

Not provided

Brand: Medtronic Neuromodulation. Contact: Medtronic Neuromodulation.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-2118-2014

Event ID

68615

Units affected

300

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 6 Aug 2014. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button