Open official notice

LINK MP Monoblock Hip Stem, Trial Stem Instruments, Size 14-Indicated for patients with Mobility-limiting diseases, fractures or defects of

Official notice

16 Aug 2023

RecallScope risk estimate:MediumFDA

An official FDA notice concerns LINK MP Monoblock Hip Stem, Trial Stem Instruments, Size 14-Indicated for patients with Mobility-limiting diseases, fractures or defects of. Increased force may be required to remove the LINK MP Monoblock trial hip stem (Sizes 14mm - 25mm),after impaction. Potential for prolongation of surgery or the surgical procedure may have to be modified due to this issue Published 2023-08-16.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

14

Recall number

Z-2345-2023

Official notice

2023-08-16

The hazard

Increased force may be required to remove the LINK MP Monoblock trial hip stem (Sizes 14mm - 25mm),after impaction. Potential for prolongation of surgery or the surgical procedure may have to be modified due to this issue

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

B923005

Other

UDI-DI 04026575182084

Event Ids

92758

Recall Numbers

Z-2345-2023

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

LINK MP Monoblock Hip Stem, Trial Stem Instruments, Size 14-Indicated for patients with Mobility-limiting diseases, fractures or defects of the hip joint or proximal femur. Item Number: 136-114/00

Units affected

14

Hazard

Increased force may be required to remove the LINK MP Monoblock trial hip stem (Sizes 14mm - 25mm),after impaction. Potential for prolongation of surgery or the surgical procedure may have to be modified due to this issue

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Company & contacts

Manufacturer

Waldemar Link GmbH & Co. KG (Mfg Site)

Phone

Not provided

Brand: Waldemar Link Kg. Contact: Waldemar Link GmbH & Co. KG (Mfg Site).

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-2345-2023

Event ID

92758

Units affected

14

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 16 Aug 2023. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button