LeadCare II Blood Lead Test Kit Catalog Number 70-6762

Official notice

7 Jul 2021

RecallScope risk estimate:CriticalFDA

An official FDA notice concerns LeadCare II Blood Lead Test Kit Catalog Number 70-6762. Magellan has received reports that control tests of either the Low-Control and/or the High-Control generated a low result. The issue may cause the underestimation of blood lead levels in patient samples. Published 2021-07-07.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

387,281

Recall number

Z-1952-2021

Official notice

2021-07-07

The hazard

Magellan has received reports that control tests of either the Low-Control and/or the High-Control generated a low result. The issue may cause the underestimation of blood lead levels in patient samples.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

87992

Recall Numbers

Z-1952-2021

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

LeadCare II Blood Lead Test Kit Catalog Number 70-6762

Units affected

387,281

Hazard

Magellan has received reports that control tests of either the Low-Control and/or the High-Control generated a low result. The issue may cause the underestimation of blood lead levels in patient samples.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AU · Australia

Affected area 3

CA · Canada

Affected area 4

HK · Hong Kong

Affected area 5

IN · India

Affected area 6

IT · Italy

Affected area 7

NZ · New Zealand

Affected area 8

PE · Peru

Affected area 9

VN · Vietnam

Company & contacts

Manufacturer

Magellan Diagnostics, Inc.

Phone

Not provided

Brand: Magellan Diagnostics. Contact: Magellan Diagnostics, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class I

Recall details

Recall number

Z-1952-2021

Event ID

87992

Units affected

387,281

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 7 Jul 2021. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button