Open official notice

Lag screw 3.2mm guide pin sleeve, ref 71674032 - product usage: the trigen intertan nails are indicated for fractures of the femur

Official notice

20 Nov 2019

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Lag screw 3.2mm guide pin sleeve, ref 71674032 - product usage: the trigen intertan nails are indicated for fractures of the femur. LAG SCREW 3.2 GUIDE PIN SLEEVE are being recalled due to a manufacturing error. The detent balls in the guide pin sleeve were not drilled to the correct depth, which will cause them to not fully compress. Published 2019-11-20.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

116

Recall number

Z-0351-2020

Official notice

2019-11-20

The hazard

LAG SCREW 3.2 GUIDE PIN SLEEVE are being recalled due to a manufacturing error. The detent balls in the guide pin sleeve were not drilled to the correct depth, which will cause them to not fully compress.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

84059

Recall Numbers

Z-0351-2020

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

LAG SCREW 3.2MM GUIDE PIN SLEEVE, REF 71674032 - Product Usage: The TRIGEN InterTAN nails are indicated for fractures of the femur including: simple shaft fractures, comminuted shaft fractures, spiral shaft fractures, long oblique shaft fractures and segmental shaft fractures; subtrochanteric fractures; intertrochanteric fractures; ipsilateral femoral shaft/neck fractures; intracapsular fractures; nonunions and malunions; polytrauma and multiple fractures; prophylactic nailing of impending pathologic fractures; reconstruction, following tumor resection and grafting; bone lengthening and shortening.

Units affected

116

Hazard

LAG SCREW 3.2 GUIDE PIN SLEEVE are being recalled due to a manufacturing error. The detent balls in the guide pin sleeve were not drilled to the correct depth, which will cause them to not fully compress.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AE · United Arab Emirates

Affected area 3

BE · Belgium

Affected area 4

BR · Brazil

Affected area 5

CN · China

Affected area 6

ES · Spain

Affected area 7

FR · France

Affected area 8

GB · United Kingdom

Affected area 9

IN · India

Affected area 10

JP · Japan

Affected area 11

RU · Russia

Affected area 12

SG · Singapore

Affected area 13

ZA · South Africa

Company & contacts

Manufacturer

Smith & Nephew, Inc.

Phone

Not provided

Brand: Smith Nephew. Contact: Smith & Nephew, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0351-2020

Event ID

84059

Units affected

116

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 20 Nov 2019. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button