Open official notice

Intelepacs, inteleconnect portal tech portal versions pacs-5-1-1-r102, pacs-5-2-1-r93, pacs-5-3-1-r83 through r128, pacs-5-4-1-r66 through

Official notice

21 Oct 2026

Last checked

26 Oct 2026

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Intelepacs, inteleconnect portal tech portal versions pacs-5-1-1-r102, pacs-5-2-1-r93, pacs-5-3-1-r83 through r128, pacs-5-4-1-r66 through. Following a Chromium-based web-browser, report and displayed images may not match. Published 2026-10-21.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

317

Recall number

Z-0186-2027

Official notice

2026-10-21

The hazard

Following a Chromium-based web-browser, report and displayed images may not match.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

99899

Recall Numbers

Z-0186-2027

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

IntelePACS, InteleConnect Portal/Tech Portal Versions PACS-5-1-1-R102, PACS-5-2-1-R93, PACS-5-3-1-R83 through R128, PACS-5-4-1-R66 through R101, PACS-5-5-1-R50 through R89, PACS-5-6-1-R41 through R81, PACS-5-7-1-R21 through R61, PACS-5-8-1-R2 through R42, PACS-5-9-1-R8 through R19; System, Image Processing, Radiological

Units affected

317

Hazard

Following a Chromium-based web-browser, report and displayed images may not match.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Company & contacts

Manufacturer

INTELERAD MEDICAL SYSTEMS INCORPORATED

Phone

Not provided

Brand: Intelerad Medical. Contact: INTELERAD MEDICAL SYSTEMS INCORPORATED.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0186-2027

Event ID

99899

Units affected

317

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 21 Oct 2026. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button