Integra Licox Brain Tissue Oxygen Monitoring Integra Licox Recon Tissue Oxygen Monitoring 1) LICOX¿ Complete Brain Probe Kit (comprised of IP1, CC1.P1) 2) LICOX¿ Complete Brain Probe Kit (comprised of IP2, CC1.P1) 3) LICOX¿ Complete Brain Tunnelling Probe Kit (comprised of VK5.2, CC1.P1 4) LICOX¿ Complete Brain Probe Kit (comprised of IM1, CC1.SB) Sterile Rx Only Manufacturer CMS -Gesse schaft for medizinische Sondentechnik mbH Dorfstrasse 2 24247 <eilkendorf Germany Distributed by Integra NeuroSciences 311 Enterprise Drive Plainsboro, NJ 08536, USA

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

2,506

Recall number

Z-0831-2015

Official notice

2014-12-24

The hazard

Due to a complaint, it was determined that multiple lots of Licox catheters manufactured since October 29, 2013 contained a smart card which was programmed with an inaccurate lkN2 value.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

69803

Recall Numbers

Z-0831-2015

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Integra Licox Brain Tissue Oxygen Monitoring Integra Licox Recon Tissue Oxygen Monitoring 1) LICOX¿ Complete Brain Probe Kit (comprised of IP1, CC1.P1) 2) LICOX¿ Complete Brain Probe Kit (comprised of IP2, CC1.P1) 3) LICOX¿ Complete Brain Tunnelling Probe Kit (comprised of VK5.2, CC1.P1 4) LICOX¿ Complete Brain Probe Kit (comprised of IM1, CC1.SB) Sterile Rx Only Manufacturer CMS -Gesse;schaft for medizinische Sondentechnik mbH Dorfstrasse 2 24247 <eilkendorf Germany Distributed by Integra NeuroSciences 311 Enterprise Drive Plainsboro, NJ 08536, USA

Units affected

2,506

Hazard

Due to a complaint, it was determined that multiple lots of Licox catheters manufactured since October 29, 2013 contained a smart card which was programmed with an inaccurate lkN2 value.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Company & contacts

Manufacturer

Integra LifeSciences Corp.

Phone

Not provided

Brand: Integra Lifesciences. Contact: Integra LifeSciences Corp..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0831-2015

Event ID

69803

Units affected

2,506

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 24 Dec 2014. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button