Open official notice

Infusomat Space Infusion System Large Volume Pump, Product Code 8713050U

Official notice

18 Sept 2024

RecallScope risk estimate:CriticalFDA

An official FDA notice concerns Infusomat Space Infusion System Large Volume Pump, Product Code 8713050U. Extension of previous recall RES 92978. Pumps have the potential for upstream occlusion sensor to cease proper function due to buildup of electrostatic charge during operation triggering false downstream and upstream occlusion alarms. Additional devices have been affected. Published 2024-09-18.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

14

Recall number

Z-2972-2024

Official notice

2024-09-18

The hazard

Extension of previous recall RES 92978. Pumps have the potential for upstream occlusion sensor to cease proper function due to buildup of electrostatic charge during operation triggering false downstream and upstream occlusion alarms. Additional devices have been affected.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Other

UDI-DI 04046963716752

Event Ids

95093

Recall Numbers

Z-2972-2024

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Infusomat Space Infusion System/ Large Volume Pump, Product Code 8713050U

Units affected

14

Hazard

Extension of previous recall RES 92978. Pumps have the potential for upstream occlusion sensor to cease proper function due to buildup of electrostatic charge during operation triggering false downstream and upstream occlusion alarms. Additional devices have been affected.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States · AL

Location 2

US · United States · CA

Location 3

US · United States · GA

Location 4

US · United States · IL

Location 5

US · United States · MI

Location 6

US · United States · TX

Affected area 1

US · United States

Company & contacts

Manufacturer

B Braun Medical Inc

Phone

Not provided

Brand: B Braun Medical. Contact: B Braun Medical Inc.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class I

Recall details

Recall number

Z-2972-2024

Event ID

95093

Units affected

14

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 18 Sept 2024. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.