Infusomat(R) Space Pump (US Version) with software version 686G030103 (commonly termed G03), 686E030003, 686F030005, 686F030007

Official notice

25 Jul 2012

RecallScope risk estimate:CriticalFDA

An official FDA notice concerns Infusomat(R) Space Pump (US Version) with software version 686G030103 (commonly termed G03), 686E030003, 686F030005, 686F030007. Firm became aware of the potential for breakage of the anti free flow clip catch, located on the inside of the pump door, when the IV set anti free flow clip is incorrectly inserted into the pump and the pump door forced closed. If the clip catch is broken and the door opened, free flow protection is still ensured. However, after the set is manually removed from the pump without the closure of t Published 2012-07-25.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

38,225

Recall number

Z-1992-2012

Official notice

2012-07-25

The hazard

Firm became aware of the potential for breakage of the anti free flow clip catch, located on the inside of the pump door, when the IV set anti free flow clip is incorrectly inserted into the pump and the pump door forced closed. If the clip catch is broken and the door opened, free flow protection is still ensured. However, after the set is manually removed from the pump without the closure of t

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Skus

8713050U

Event Ids

62115

Model numbers

8713050U

Recall Numbers

Z-1992-2012

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Infusomat(R) Space Pump (US Version) with software version 686G030103 (commonly termed 'G03'), 686E030003, 686F030005, 686F030007, 686G030002, and 686G030102. Intended for use with adults, pediatrics and neonates and is intended to provide infusions of parenteral fluids/medications, blood and blood products indicated for infusion through FDA approved routes of administration.

Units affected

38,225

Hazard

Firm became aware of the potential for breakage of the anti free flow clip catch, located on the inside of the pump door, when the IV set anti free flow clip is incorrectly inserted into the pump and the pump door forced closed. If the clip catch is broken and the door opened, free flow protection is still ensured. However, after the set is manually removed from the pump without the closure of t

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Company & contacts

Manufacturer

B. Braun Medical, Inc.

Phone

Not provided

Brand: B Braun Medical. Contact: B. Braun Medical, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class I

Recall details

Recall number

Z-1992-2012

Event ID

62115

Units affected

38,225

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 25 Jul 2012. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button