Open official notice

Infant Heel Warmer with Attachment Tab, Nonsterile, Active Product SKU MH00002T The infant heel warmer is activated at the point of care by

Official notice

28 Jul 2021

RecallScope risk estimate:HighFDA

An official FDA notice concerns Infant Heel Warmer with Attachment Tab, Nonsterile, Active Product SKU MH00002T The infant heel warmer is activated at the point of care by. A caution statement is being added to remind the user to activate away from face and infant as reports of pack rupture or leaking during activation may resulting in the contents splashing into the user s eyes or onto the infant. Potential harms associated with contact of the liquid inside the pouch are rash, burn and inflammation. Published 2021-07-28.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

926,425

Recall number

Z-2094-2021

Official notice

2021-07-28

The hazard

A caution statement is being added to remind the user to activate away from face and infant as reports of pack rupture or leaking during activation may resulting in the contents splashing into the user s eyes or onto the infant. Potential harms associated with contact of the liquid inside the pouch are rash, burn and inflammation.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

88173

Recall Numbers

Z-2094-2021

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Infant Heel Warmer with Attachment Tab, Nonsterile, Active Product SKU MH00002T The infant heel warmer is activated at the point of care by squeezing the pouch. This mixes the substrates inside the pouch causing an exothermic chemical reaction, warming the pouch, which is then applied to the heel of an infant to aid in blood draws

Units affected

926,425

Hazard

A caution statement is being added to remind the user to activate away from face and infant as reports of pack rupture or leaking during activation may resulting in the contents splashing into the user s eyes or onto the infant. Potential harms associated with contact of the liquid inside the pouch are rash, burn and inflammation.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Company & contacts

Manufacturer

Cardinal Health 200, LLC

Phone

Not provided

Brand: Cardinal Health 200. Contact: Cardinal Health 200, LLC.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-2094-2021

Event ID

88173

Units affected

926,425

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 28 Jul 2021. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button