Open official notice

Halyard anterior hip pack. Model catalog number: soah31

Official notice

28 Oct 2026

Last checked

2 Nov 2026

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Halyard anterior hip pack. Model catalog number: soah31. A non-sterile component was piggybacked and incorrectly placed outside of the sterile barrier and not subjected to EO sterilization per kit specifications. Published 2026-10-28.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

56

Recall number

Z-0282-2027

Official notice

2026-10-28

The hazard

A non-sterile component was piggybacked and incorrectly placed outside of the sterile barrier and not subjected to EO sterilization per kit specifications.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

1689215

Skus

SOAH31

Other

UDI-DI 10809160512890

Event Ids

99968

Model numbers

SOAH31

Recall Numbers

Z-0282-2027

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Halyard ANTERIOR HIP PACK. Model/Catalog Number: SOAH31.

Units affected

56

Hazard

A non-sterile component was piggybacked and incorrectly placed outside of the sterile barrier and not subjected to EO sterilization per kit specifications.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Company & contacts

Manufacturer

AVID Medical, Inc.

Phone

Not provided

Brand: Avid Medical. Contact: AVID Medical, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0282-2027

Event ID

99968

Units affected

56

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 28 Oct 2026. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button