Guardian Connect App CSS7200 iOS and Guardian Connect Transmitter GST4C used on the iPhone, iPad and iPod Touch devices

Official notice

3 Jun 2020

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Guardian Connect App CSS7200 iOS and Guardian Connect Transmitter GST4C used on the iPhone, iPad and iPod Touch devices. Customers using the firm's continuous glucose monitoring system application on an iPhone, iPad or iPod Touch with ios software version 12, 12.1, or 12,2 are likely to experience a shortened transmitter battery life (approximately 4-5 days instead of the normal 7 days or more) after a full charge. Published 2020-06-03.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

23,978

Recall number

Z-2106-2020

Official notice

2020-06-03

The hazard

Customers using the firm's continuous glucose monitoring system application on an iPhone, iPad or iPod Touch with ios software version 12, 12.1, or 12,2 are likely to experience a shortened transmitter battery life (approximately 4-5 days instead of the normal 7 days or more) after a full charge.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Skus

CSS7200

Event Ids

85091

Model numbers

CSS7200

Recall Numbers

Z-2106-2020

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Guardian Connect App CSS7200 iOS and Guardian Connect Transmitter GST4C used on the iPhone, iPad and iPod Touch devices.

Units affected

23,978

Hazard

Customers using the firm's continuous glucose monitoring system application on an iPhone, iPad or iPod Touch with ios software version 12, 12.1, or 12,2 are likely to experience a shortened transmitter battery life (approximately 4-5 days instead of the normal 7 days or more) after a full charge.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Company & contacts

Manufacturer

Medtronic Inc.

Phone

Not provided

Brand: Medtronic. Contact: Medtronic Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-2106-2020

Event ID

85091

Units affected

23,978

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 3 Jun 2020. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button