GE, Revolution CT. The system is intended for head, whole body, cardiac, and vascular X-ray Computed Tomography applications

Official notice

11 May 2016

RecallScope risk estimate:MediumFDA

An official FDA notice concerns GE, Revolution CT. The system is intended for head, whole body, cardiac, and vascular X-ray Computed Tomography applications. An out of tolerance torque tool was used on three joints during the manufacturing process. If the X-ray Tube becomes loose on the CT gantry during operation, this could result in beam tracking errors, potential scan aborts and/or a possible image artifact. There is a secondary safety mechanism to prevent components from being expelled from the gantry if the bolts fail. Published 2016-05-11.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

26

Recall number

Z-1589-2016

Official notice

2016-05-11

The hazard

An out of tolerance torque tool was used on three joints during the manufacturing process. If the X-ray Tube becomes loose on the CT gantry during operation, this could result in beam tracking errors, potential scan aborts and/or a possible image artifact. There is a secondary safety mechanism to prevent components from being expelled from the gantry if the bolts fail.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

73974

Recall Numbers

Z-1589-2016

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

GE, Revolution CT. The system is intended for head, whole body, cardiac, and vascular X-ray Computed Tomography applications

Units affected

26

Hazard

An out of tolerance torque tool was used on three joints during the manufacturing process. If the X-ray Tube becomes loose on the CT gantry during operation, this could result in beam tracking errors, potential scan aborts and/or a possible image artifact. There is a secondary safety mechanism to prevent components from being expelled from the gantry if the bolts fail.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States · FL

Location 2

US · United States · GA

Location 3

US · United States · NC

Location 4

US · United States · TX

Affected area 1

US · United States

Affected area 2

AE · United Arab Emirates

Affected area 3

AU · Australia

Affected area 4

CA · Canada

Affected area 5

CN · China

Affected area 6

DE · Germany

Affected area 7

FR · France

Affected area 8

GB · United Kingdom

Affected area 9

IT · Italy

Affected area 10

JP · Japan

Affected area 11

KR · South Korea

Affected area 12

NO · Norway

Affected area 13

PE · Peru

Affected area 14

SE · Sweden

Company & contacts

Manufacturer

GE Medical Systems, LLC

Phone

Not provided

Brand: GE Medical. Contact: GE Medical Systems, LLC.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1589-2016

Event ID

73974

Units affected

26

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 11 May 2016. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button