Open official notice

For intraventricular indications mainly used for operations within the brain ventricles. Product Code: FF399R

Official notice

13 Sept 2023

RecallScope risk estimate:MediumFDA

An official FDA notice concerns For intraventricular indications mainly used for operations within the brain ventricles. Product Code: FF399R. Trocar manufactured with the shaft too long and does not meet manufacturing specifications. As a result, when used with the 30¿ neuroendoscope, a portion of the visual field is obscured. Published 2023-09-13.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

25

Recall number

Z-2518-2023

Official notice

2023-09-13

The hazard

Trocar manufactured with the shaft too long and does not meet manufacturing specifications. As a result, when used with the 30¿ neuroendoscope, a portion of the visual field is obscured.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

52750146

Other

UDI-DI 04038653065547

Event Ids

92813

Recall Numbers

Z-2518-2023

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

For intraventricular indications mainly used for operations within the brain ventricles. Product Code: FF399R

Units affected

25

Hazard

Trocar manufactured with the shaft too long and does not meet manufacturing specifications. As a result, when used with the 30¿ neuroendoscope, a portion of the visual field is obscured.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Company & contacts

Manufacturer

Aesculap Implant Systems LLC

Phone

Not provided

Brand: Aesculap Implant. Contact: Aesculap Implant Systems LLC.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-2518-2023

Event ID

92813

Units affected

25

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 13 Sept 2023. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button