Open official notice

Flex. Grasp. Forceps 5fr wl 550mm, product number 828.051

Official notice

3 May 2023

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Flex. Grasp. Forceps 5fr wl 550mm, product number 828.051. Potentially unable to open forceps jaws when the endoscope shaft is bent by about 90 degrees. If, in addition, the proximal forceps shaft is bent, the forceps can no longer be opened or restrictions begin even with a slight bending of the endoscope. Published 2023-05-03.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

838

Recall number

Z-1422-2023

Official notice

2023-05-03

The hazard

Potentially unable to open forceps jaws when the endoscope shaft is bent by about 90 degrees. If, in addition, the proximal forceps shaft is bent, the forceps can no longer be opened or restrictions begin even with a slight bending of the endoscope.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

4500377299

Event Ids

91953

Recall Numbers

Z-1422-2023

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

FLEX. GRASP. FORCEPS 5FR WL 550MM, Product Number 828.051

Units affected

838

Hazard

Potentially unable to open forceps jaws when the endoscope shaft is bent by about 90 degrees. If, in addition, the proximal forceps shaft is bent, the forceps can no longer be opened or restrictions begin even with a slight bending of the endoscope.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AE · United Arab Emirates

Affected area 3

AR · Argentina

Affected area 4

AT · Austria

Affected area 5

BE · Belgium

Affected area 6

BR · Brazil

Affected area 7

CA · Canada

Affected area 8

CL · Chile

Affected area 9

CZ · Czech Republic

Affected area 10

DE · Germany

Affected area 11

DK · Denmark

Affected area 12

ES · Spain

Affected area 13

FI · Finland

Affected area 14

FR · France

Affected area 15

GR · Greece

Affected area 16

HK · Hong Kong

Affected area 17

HR · Croatia

Affected area 18

IE · Ireland

Affected area 19

IN · India

Affected area 20

IT · Italy

Affected area 21

JO · Jordan

Affected area 22

JP · Japan

Affected area 23

KR · South Korea

Affected area 24

KW · Kuwait

Affected area 25

LT · Lithuania

Affected area 26

MX · Mexico

Affected area 27

NL · Netherlands

Affected area 28

NO · Norway

Affected area 29

NZ · New Zealand

Affected area 30

PK · Pakistan

Affected area 31

PL · Poland

Affected area 32

PT · Portugal

Affected area 33

RO · Romania

Affected area 34

RS · Serbia

Affected area 35

SA · Saudi Arabia

Affected area 36

SI · Slovenia

Affected area 37

SK · Slovakia

Affected area 38

TH · Thailand

Affected area 39

TR · Turkey

Company & contacts

Manufacturer

Richard Wolf GmbH

Phone

Not provided

Brand: Richard Wolf. Contact: Richard Wolf GmbH.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1422-2023

Event ID

91953

Units affected

838

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 3 May 2023. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button