Open official notice

Fellows Mighty 8 Power strip (Part Number 04175-0008) and TrippLite Power strip (PS-415-HGULTRA - Part Number 13-00106-00) are two

Official notice

24 Jan 2018

RecallScope risk estimate:MediumFDA

Fellows "Mighty 8" Power strip (Part Number 04175-0008) and TrippLite Power strip (PS-415-HGULTRA - Part Number 13-00106-00) are two components sold as part of the Merge Hemo System.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

1,150

Recall number

Z-0350-2018

Official notice

2018-01-24

The hazard

Merge is conducting a hardware recall (replacement) of Fellows "Mighty 8" Power Strip and TrippLite Power Strip due to overheat of input resistor causing the power strip to smoke.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

76847

Recall Numbers

Z-0350-2018

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Fellows "Mighty 8" Power strip (Part Number 04175-0008) and TrippLite Power strip (PS-415-HGULTRA - Part Number 13-00106-00) are two components sold as part of the Merge Hemo System.

Units affected

1,150

Hazard

Merge is conducting a hardware recall (replacement) of Fellows "Mighty 8" Power Strip and TrippLite Power Strip due to overheat of input resistor causing the power strip to smoke.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Company & contacts

Manufacturer

Merge Healthcare, Inc.

Phone

Not provided

Brand: Merge Healthcare. Contact: Merge Healthcare, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0350-2018

Event ID

76847

Units affected

1,150

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 24 Jan 2018. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button