EVair air compressor, model numbers: M1230849 and M1230847

Official notice

7 Feb 2024

RecallScope risk estimate:CriticalFDA

An official FDA notice concerns EVair air compressor, model numbers: M1230849 and M1230847. GE HealthCare has become aware of elevated levels of formaldehyde when the EVair or EVair 03 (Jun-Air) optional compressors are used with the CARESCAPE R860 or Engstr¿m Carestation/Pro ventilators, respectively. These elevated results were observed in preliminary testing that was conducted at an elevated room temperature of 40oC (104oF), at the lowest possible flow condition of 2 L/min (worstcase, minimum bias flow with no additional ventilation), and all of the gas being supplied from the compressor (i.e., FiO2 of 21% / no supplemental oxygen). GE HealthCare is continuing to evaluate the root cause for these elevated levels of formaldehyde. While these test conditions are not representative of typical clinical use conditions. GE HealthCare is taking this action to further reduce the potential for patient exposure. Published 2024-02-07.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

4,032

Recall number

Z-0774-2024

Official notice

2024-02-07

The hazard

GE HealthCare has become aware of elevated levels of formaldehyde when the EVair or EVair 03 (Jun-Air) optional compressors are used with the CARESCAPE R860 or Engstr¿m Carestation/Pro ventilators, respectively. These elevated results were observed in preliminary testing that was conducted at an elevated room temperature of 40oC (104oF), at the lowest possible flow condition of 2 L/min (worstcase, minimum bias flow with no additional ventilation), and all of the gas being supplied from the compressor (i.e., FiO2 of 21% / no supplemental oxygen). GE HealthCare is continuing to evaluate the root cause for these elevated levels of formaldehyde. While these test conditions are not representative of typical clinical use conditions. GE HealthCare is taking this action to further reduce the potential for patient exposure.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

93715

Recall Numbers

Z-0774-2024

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

EVair air compressor, model numbers: M1230849 and M1230847

Units affected

4,032

Hazard

GE HealthCare has become aware of elevated levels of formaldehyde when the EVair or EVair 03 (Jun-Air) optional compressors are used with the CARESCAPE R860 or Engstr¿m Carestation/Pro ventilators, respectively. These elevated results were observed in preliminary testing that was conducted at an elevated room temperature of 40oC (104oF), at the lowest possible flow condition of 2 L/min (worstcase, minimum bias flow with no additional ventilation), and all of the gas being supplied from the compressor (i.e., FiO2 of 21% / no supplemental oxygen). GE HealthCare is continuing to evaluate the root cause for these elevated levels of formaldehyde. While these test conditions are not representative of typical clinical use conditions. GE HealthCare is taking this action to further reduce the potential for patient exposure.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Company & contacts

Manufacturer

Datex-Ohmeda, Inc.

Phone

Not provided

Brand: Datex-ohmeda. Contact: Datex-Ohmeda, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class I

Recall details

Recall number

Z-0774-2024

Event ID

93715

Units affected

4,032

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 7 Feb 2024. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button