Open official notice

Endoscopic retrieval device. The disposable Raptor grasping device in intended to retrieve foreign bodies or resected tissue specimens in

Official notice

10 Oct 2012

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Endoscopic retrieval device. The disposable Raptor grasping device in intended to retrieve foreign bodies or resected tissue specimens in. When the outer catheter is either coiled or in a contorted configuration outside of the endoscope and the patient, and excessive force is applied to the handle, the jaws may become inoperable and may become stuck in the open condition. Published 2012-10-10.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

1,153

Recall number

Z-0006-2013

Official notice

2012-10-10

The hazard

When the outer catheter is either coiled or in a contorted configuration outside of the endoscope and the patient, and excessive force is applied to the handle, the jaws may become inoperable and may become stuck in the open condition.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

63196

Recall Numbers

Z-0006-2013

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Endoscopic retrieval device. The disposable Raptor grasping device in intended to retrieve foreign bodies or resected tissue specimens in endoscopic procedures of the upper and lower gastrointestinal tract or biliary tract.

Units affected

1,153

Hazard

When the outer catheter is either coiled or in a contorted configuration outside of the endoscope and the patient, and excessive force is applied to the handle, the jaws may become inoperable and may become stuck in the open condition.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AE · United Arab Emirates

Affected area 3

AR · Argentina

Affected area 4

AT · Austria

Affected area 5

AU · Australia

Affected area 6

BE · Belgium

Affected area 7

BR · Brazil

Affected area 8

CH · Switzerland

Affected area 9

CL · Chile

Affected area 10

CZ · Czech Republic

Affected area 11

DE · Germany

Affected area 12

DK · Denmark

Affected area 13

EE · Estonia

Affected area 14

ES · Spain

Affected area 15

FI · Finland

Affected area 16

FR · France

Affected area 17

GB · United Kingdom

Affected area 18

GR · Greece

Affected area 19

HR · Croatia

Affected area 20

IE · Ireland

Affected area 21

IL · Israel

Affected area 22

IN · India

Affected area 23

IT · Italy

Affected area 24

JO · Jordan

Affected area 25

LT · Lithuania

Affected area 26

MT · Malta

Affected area 27

MX · Mexico

Affected area 28

NL · Netherlands

Affected area 29

NO · Norway

Affected area 30

NZ · New Zealand

Affected area 31

PL · Poland

Affected area 32

PT · Portugal

Affected area 33

RO · Romania

Affected area 34

SA · Saudi Arabia

Affected area 35

SI · Slovenia

Affected area 36

SV · El Salvador

Affected area 37

TR · Turkey

Affected area 38

ZA · South Africa

Company & contacts

Manufacturer

US Endoscopy Group Inc

Phone

Not provided

Brand: Endoscopy. Contact: US Endoscopy Group Inc.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0006-2013

Event ID

63196

Units affected

1,153

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 10 Oct 2012. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button