Image not available in RecallScope.
Empowr PS Knee System Box Cut Guide; Used to make the housing resection for the post of the posterior stabilized implant
Official notice
9 Nov 2016
Empowr PS Knee System Box Cut Guide; Used to make the housing resection for the post of the posterior stabilized implant.
What to do now
This recall has been terminated. See the official FDA notice for the latest status.
Units affected
436
Recall number
Z-0325-2017
Official notice
2016-11-09
The hazard
The hazard
It has been discovered that during Empowr PS Total Knee replacement, while chiseling for bone removal in the box area, femoral chipping has occurred in surgery. The current design of the captured box cut guides can create an instance where the surgeon is not fully completing the box cut on the side.
How to check if you’re affected
How to check if you’re affected
Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.
Event Ids
Recall Numbers
Checked manually this time?
Follow brands and keywords in the app to receive alerts for matching official notices.
Turn on recall alertsAffected products
Affected products
Product 1
Units affected
Hazard
Where it was sold
Where it was sold
Retailer information is not currently available in RecallScope. Check the official notice for distribution details.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand: Encore Medical Lp. Contact: Encore Medical, Lp.
Source & status
Source & status
Recall details
Recall details
Recall number
Event ID
Units affected
Status
Source tier
Documents, notices & source