Open official notice

Custom ti mak rs oss segmental femoral rotating hinged knee. This rotating hinged knee is intended to replace patient anatomy

Official notice

15 Nov 2017

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Custom ti mak rs oss segmental femoral rotating hinged knee. This rotating hinged knee is intended to replace patient anatomy. A custom implant manufactured entirely from titanium for a patient with a nickel allergy was requested by a surgeon. After the device was manufactured and used in the planned surgery, it was discovered internally that the implant assembly had a CoCrMo screw included as a component of the implant, which contains nickel. Published 2017-11-15.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

1

Recall number

Z-0070-2018

Official notice

2017-11-15

The hazard

A custom implant manufactured entirely from titanium for a patient with a nickel allergy was requested by a surgeon. After the device was manufactured and used in the planned surgery, it was discovered internally that the implant assembly had a CoCrMo screw included as a component of the implant, which contains nickel.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

073840

Event Ids

78299

Recall Numbers

Z-0070-2018

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Custom Ti MAK RS OSS Segmental Femoral Rotating Hinged Knee. This Rotating Hinged Knee is intended to replace patient anatomy.

Units affected

1

Hazard

A custom implant manufactured entirely from titanium for a patient with a nickel allergy was requested by a surgeon. After the device was manufactured and used in the planned surgery, it was discovered internally that the implant assembly had a CoCrMo screw included as a component of the implant, which contains nickel.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Company & contacts

Manufacturer

Zimmer Biomet, Inc.

Phone

Not provided

Brand: Zimmer Biomet. Contact: Zimmer Biomet, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0070-2018

Event ID

78299

Units affected

1

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 15 Nov 2017. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button