Crit Line in a Clip (CLiC). CliC P N: CL 10041001 CLiC is a continuous real-time monitor for non-invasive hematocrit oxygen saturation and

Official notice

29 Jun 2016

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Crit Line in a Clip (CLiC). CliC P N: CL 10041001 CLiC is a continuous real-time monitor for non-invasive hematocrit oxygen saturation and. Use of the Blood Pressure Graph located on the Crit-Line screen of the 2008T Hemodialysis Machine may result in a frozen screen, possibly causing the blood pump to stop. Published 2016-06-29.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

875

Recall number

Z-2044-2016

Official notice

2016-06-29

The hazard

Use of the Blood Pressure Graph located on the Crit-Line screen of the 2008T Hemodialysis Machine may result in a frozen screen, possibly causing the blood pump to stop.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

74241

Recall Numbers

Z-2044-2016

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Crit Line in a Clip (CLiC). CliC P/N: CL 10041001 CLiC is a continuous real-time monitor for non-invasive hematocrit oxygen saturation and percent change in blood volume measurement during hemodialysis treatment. Used in conjunction with the Fresenius 2008T Hemodialysis Machines as follows: Model: 190895 2008T GEN 2 BIBAG w/o CDX 190713 2008T Hemodialysis SYS with CDX 190858 2008T Hemodialysis System w/o CDX 190766 2008T Hemodialysis System w/o CDX

Units affected

875

Hazard

Use of the Blood Pressure Graph located on the Crit-Line screen of the 2008T Hemodialysis Machine may result in a frozen screen, possibly causing the blood pump to stop.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Company & contacts

Manufacturer

Fresenius Medical Care Renal Therapies Group, LLC

Phone

Not provided

Brand: Fresenius Medical Care Renal Therapies. Contact: Fresenius Medical Care Renal Therapies Group, LLC.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-2044-2016

Event ID

74241

Units affected

875

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 29 Jun 2016. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button