Open official notice

Coonmad morrey total elbow elbow component replacement set bushings & pins for use w size regular elbow elbow comp replacement-reg set

Official notice

26 Dec 2012

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Coonmad morrey total elbow elbow component replacement set bushings & pins for use w size regular elbow elbow comp replacement-reg set. Zimmer is initiating a lot specific recall because affected lots could include an extra Small Inner Pin instead of the required Regular Inner Pin in the Coonrad/Morrey Elbow Replacement Set and the New Ulnar Revision Kit . Published 2012-12-26.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

23

Recall number

Z-0537-2013

Official notice

2012-12-26

The hazard

Zimmer is initiating a lot specific recall because affected lots could include an extra Small Inner Pin instead of the required Regular Inner Pin in the Coonrad/Morrey Elbow Replacement Set and the New Ulnar Revision Kit .

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

61644802

Event Ids

63641

Recall Numbers

Z-0537-2013

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

COONMAD/MORREY TOTAL ELBOW ELBOW COMPONENT REPLACEMENT SET BUSHINGS & PINS FOR USE W/SIZE REGULAR ELBOW ELBOW COMP REPLACEMENT-REG SET STERILE PROD CLASS: 4100 OPERATION:9000 QTY/PKG:1 Product Usage: The ulnar and humeral components are connected by placing the hollow, outer axis pin across the two components and securing It with the solid Internal axis pin. The Inner and outer pin connection Is articulation point of the placement elbow.

Units affected

23

Hazard

Zimmer is initiating a lot specific recall because affected lots could include an extra Small Inner Pin instead of the required Regular Inner Pin in the Coonrad/Morrey Elbow Replacement Set and the New Ulnar Revision Kit .

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

CA · Canada

Affected area 3

CH · Switzerland

Affected area 4

DE · Germany

Affected area 5

FI · Finland

Affected area 6

GB · United Kingdom

Affected area 7

IT · Italy

Affected area 8

JP · Japan

Company & contacts

Brand

Manufacturer

Zimmer, Inc.

Phone

Not provided

Brand: Zimmer. Contact: Zimmer, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0537-2013

Event ID

63641

Units affected

23

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 26 Dec 2012. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button