CareLink Personal Software Therapy Management Software Tool for Diabetes CareLink System Therapy Management Software Tool for Diabetes

Official notice

20 Jan 2021

RecallScope risk estimate:MediumFDA

An official FDA notice concerns CareLink Personal Software Therapy Management Software Tool for Diabetes CareLink System Therapy Management Software Tool for Diabetes. Incomplete basal rate information may be transmitted leading to a potential for inappropriate change in therapy. During data upload, radio frequency (RF) interference can occur at any point from surrounding devices or equipment that may affect the data transfer between the insulin pump and the CareLink Personal or CareLink System Software. Published 2021-01-20.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

46,461

Recall number

Z-0843-2021

Official notice

2021-01-20

The hazard

Incomplete basal rate information may be transmitted leading to a potential for inappropriate change in therapy. During data upload, radio frequency (RF) interference can occur at any point from surrounding devices or equipment that may affect the data transfer between the insulin pump and the CareLink Personal or CareLink System Software.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

86895

Recall Numbers

Z-0843-2021

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

CareLink Personal Software Therapy Management Software Tool for Diabetes CareLink System Therapy Management Software Tool for Diabetes Software Uploader versions: 3.1.5.000 3.1.2.000

Units affected

46,461

Hazard

Incomplete basal rate information may be transmitted leading to a potential for inappropriate change in therapy. During data upload, radio frequency (RF) interference can occur at any point from surrounding devices or equipment that may affect the data transfer between the insulin pump and the CareLink Personal or CareLink System Software.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AU · Australia

Affected area 3

CA · Canada

Affected area 4

CN · China

Affected area 5

IN · India

Affected area 6

JP · Japan

Affected area 7

KR · South Korea

Affected area 8

NZ · New Zealand

Company & contacts

Manufacturer

Medtronic Inc.

Phone

Not provided

Brand: Medtronic. Contact: Medtronic Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0843-2021

Event ID

86895

Units affected

46,461

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 20 Jan 2021. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button