Open official notice

CareFusion Safe-T-Centesis Catheter Drainage Tray, Sterile, For Single Use Only, Rx only, Catalog Number PIG1280T, UDI Number

Official notice

6 May 2020

RecallScope risk estimate:MediumFDA

An official FDA notice concerns CareFusion Safe-T-Centesis Catheter Drainage Tray, Sterile, For Single Use Only, Rx only, Catalog Number PIG1280T, UDI Number. The 8 Fr catheter was manufactured with the 6.9 Fr needle instead of the 8 Fr needle. Published 2020-05-06.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

900

Recall number

Z-1795-2020

Official notice

2020-05-06

The hazard

The 8 Fr catheter was manufactured with the 6.9 Fr needle instead of the 8 Fr needle.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

0001304986

Event Ids

85380

Recall Numbers

Z-1795-2020

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

CareFusion Safe-T-Centesis Catheter Drainage Tray, Sterile, For Single Use Only, Rx only, Catalog Number PIG1280T, UDI Number 01)10885403108402(17) 210930(10)0001304986

Units affected

900

Hazard

The 8 Fr catheter was manufactured with the 6.9 Fr needle instead of the 8 Fr needle.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States · AL

Location 2

US · United States · CO

Location 3

US · United States · IL

Location 4

US · United States · MA

Location 5

US · United States · MO

Location 6

US · United States · OK

Location 7

US · United States · RI

Location 8

US · United States · VA

Location 9

US · United States · WI

Affected area 1

US · United States

Company & contacts

Manufacturer

Becton Dickinson & Company

Phone

Not provided

Brand: Becton Dickinson. Contact: Becton Dickinson & Company.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1795-2020

Event ID

85380

Units affected

900

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 6 May 2020. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button